46 Auto Ref/Keratometer HRK-8000A ------------------------------------------------------------------------------------------------
Ten thousand days When Buffer Full output with the message,
DPC compensation will exceed the number of possible Dead
Pixel.
D. C in the same way and then Setup and then to the Dead Pixel.
1) KER CCD's White Pixel
2) KER CCD
‟s Black Pixel in
3) COLOR CCD's White Pixel
4) COLOR CCD
‟s Black Pixel in
E. After DPC Setup, engineer mode by pressing the SAVE DATA is
stored.
F. HRK-8000A restarts, and were being corrected by the camera
from the screen Dead Pixel Check.
4-1-12. How to upgrade OS program
(1) Turn off HRK-8000A.
(2) S
tart DNW program, set your PC‟s COM port number, 115200 Bps.
(3) Connect PC to HRK-8000A with serial cable.
(4) Pressing the 5th key, then turn on HRK-8000A
.