How to Set 12 standard lenses
If the measured value is not accurate, you should set the diopter variables again by using
the standard lens-set: for all the 12 lenses in the standard lens set. Always keep the
standard lens clean out of dust or stain.
A.
Set the diopter display step into 0.01.
B.
Place the standard lens and move it so that ‘Green Color Cross Line’ will appear.
C.
Press the ‘LENS’ (Key 1) button.
D.
Press the ‘MEASURE’ button.
E.
Determine the lens to measure by using two ‘-INDEX’, ’+INDEX’ buttons.
F.
‘D-0.01’ button and ‘D+0.01’ button by the given value in the Standard Lens-set
specification.
G.
To save and exit, press the ‘SAVE’ button.
H.
Press the ‘EXIT’ button again. And check if the diopter that has been set up in the
basic display screen is shown. .
I.
Repeat the procedures from step B to H for 12 standard lenses.
The information displayed in the ‘Setting Diopter’ screen is the same as that of ‘Setting
0D’ screen.
Summary of Contents for HLM-1
Page 1: ......
Page 2: ...Auto Lensmeter HLM 1 1 Service Manual Auto Lensmeter HLM 1...
Page 21: ...20 AUTO LENSMETER 3 Repair Standard 3 1 Removing Cover Assembly P I c t u r e...
Page 23: ...22 AUTO LENSMETER 3 2 Frame Assemblely P I c t u r e...
Page 25: ...24 AUTO LENSMETER 3 3 Back Bone Assembly P I c t u r e...
Page 27: ...26 AUTO LENSMETER 3 3 1 PD Bar Sub Assembly P I c t u r e...
Page 29: ...28 AUTO LENSMETER 3 3 2 Foot Frame Sub Assembly P I c t u r e...
Page 31: ...30 AUTO LENSMETER 3 3 3 PEN Frame Sub Assembly P I c t u r e...
Page 33: ...32 AUTO LENSMETER 3 3 4 PEN Slide Sub Assembly P I c t u r e...
Page 35: ...34 AUTO LENSMETER 3 3 5 PD Pinion Sub Assembly P I c t u r e...
Page 37: ...36 AUTO LENSMETER 3 3 6 LED Bone Sub Assembly P I c t u r e...
Page 39: ...38 AUTO LENSMETER 3 4 LCD Assembly P I c t u r e...
Page 41: ...40 AUTO LENSMETER 3 5 Under Cover Assembly P I c t u r e...
Page 44: ...Auto Lensmeter HLM 1 43...
Page 49: ...48 AUTO LENSMETER C Set the connected COM port and baud rate 115200 D Press Connect button...