4-1-4.How to check REF Setup Data.
[Purpose]
To check setup data of REF model eye currently stored in CRK-8800. CRK-8800
has two channels of LOW and HIGH diopters for REF model eye setup. LOW
diopters include 8 setup data of model eyes (-20D, -15D, -10D, -5D, -2.5D, 0D,
+2.5D, +5D), HIGH diopters include 6 setup data of model eyes (0D, +2.5D, +5D,
+10D, +15D, +20D), and there is totally 14-setup datum.
[Procedure]
(1) Enter REF SETUP DATA in Engineer Setup Mode.
(2) GRID_W and GRID_H mean the average width and height offset between
REF measure signals. SPH_C and CYL_C mean the amount of spherical and
cylindrical measure of the model eye. DIOPT mean the model eye’s setup
diopter.
(3) Press PAGE button to switch between LOW and HIGH diopters.