background image

PRODUCT

Auto Ref/Keratomet

er

SERVICE MANUAL

DOC No.

SCS-I-7RK-005

 PREPARE   REVIEW

CONFIRM

DATE

2015.11.17

MODEL

CR/CRK-7000

REVISION DATE

REVISON

 

Adjust CCD camera

USE

2015.11.17

Step 1. Entering engineer mode

Turn on Unit(CRK-7000) Power
After initial, Press button in order of 4→3→4→3→6→6
Using button 4 and 5, select [4. Motor test & setup]
Push [SEL] to enter. (button 6)

1

2

3

4

5

6

Step 2. Choose “Alignment”

Select “1.Alignment” content and press 

ENTER

 button(6

th

 button).

Step 3. Put “REF model Eye” on the head rest

     

Firmly fix the KEF Setup tool with screws on the chin rest. 

 

Summary of Contents for CR-7000

Page 1: ...RK 005 PREPARE REVIEW CONFIRM DATE 2015 11 17 MODEL CR CRK 7000 REVISION DATE REVISON Adjust CCD camera USE 2015 11 17 Fault Faulty part Cause 1 After replacing the CCD camera without 6 points 1 CCD camera position Action What When 1 Adjust CCDC camera position ...

Page 2: ...Step 1 Entering engineer mode Turn on Unit CRK 7000 Power After initial Press button in order of 4 3 4 3 6 6 Using button 4 and 5 select 4 Motor test setup Push SEL to enter button 6 1 2 3 4 5 6 Step 2 Choose Alignment Select 1 Alignment content and press ENTER button 6th button Step 3 Put REF model Eye on the head rest Firmly fix the KEF Setup tool with screws on the chin rest ...

Page 3: ...Choose MODE 1 Select MODE content and press MODE button 2nd button Appearance of Figure 2 2 Moving machine head to find the most clear position The position which the rings looks most clearly is focusing position 3 Fixed machine head Figure 1 Figure 2 4 Adjust the position of the C CD camera until the picture is s hown in Figure 3 Appear 6 point Press 6th button adjust LED 50 Figure 3 ...

Page 4: ...4 Do not need to tighten so that the CCD camera can move When you move the CCD camera pay attention to the C and Y value on the screen Other values can be ignored Adjust the camera process must be kept in the most clear position After the completion of save Figure 4 C 0 or C 1 Y 240 1 Note C and Y values must be specified otherwise it will cause CYL measure errors CCD camera position adjustment is...

Reviews: