3-3-2 Incorrect initial setting
z
Chart or Mask keeps scrolling
.
z
Halogen lamp does not light
Initial setting sequence:
1.
Chart zero detection
2. Mask zero detection
3. Halogen lamp lighting
CHECK POINTS
- In the case of abnormal zero detection(to be checked for by connecting a oscilloscope).
At this time, GND should be connected to GND. < See the pin map of Fig E1 >
Abnormal
Zero detection
Check point
(Main PCB)
Correct output
Correct action
Chart PI
J5
5V
Output available: Main board is
malfunctioning
and
should
be
replaced.
Output unavailable: Chart PI ASSY is
malfunctioning
and
should
be
replaced.
Mask PI
J7
5V
Output available: Main board is
malfunctioning
and
should
be
replaced.
Output unavailable: Large Mask PI
ASSY is malfunctioning and should be
replaced.
- In the case of that halogen lamp does not light:
1. Check the Fuse in main PCB. -> If the Fuse is off, replace the fuse.
2. Check the voltage between pin1 and pin2 of J12.
If Voltage level is 6V/AC(
±
1V/AC), then replace the halogen lamp. If voltage level is very low,
replace main PCB.