Sonicaid FM800 Encore Series Instructions for Use
29
4
3B
Setup
4.1
44B
Overview
The Setup procedures are much simpler if the differences between Setup, Current Alarms,
Default Alarms and System Settings are understood.
Setup
User preferences available directly from the Setup screen.
Current Alarms
Allows alarms to be set before connecting transducers. Settings
remain in force until FM800E is switched off. When switched on
again, the FM800E reverts to the default alarms.
Alarms can be set for:
FHR, MECG, NBP, MSpO
2
Alarms can also be set from the main screen by pressing the key
beside the parameter it is desired to change.
Default Alarms
Default alarm settings protected by the access code (2755):
Changes made using Default Alarms remain in force when FM800E is
switched off and on again.
System Settings
User preferences protected by the access code (2755):
Changes made using System Settings remain in force when FM800E
is switched off and on again.
4.2
45B
Current alarms
This option allows the alarms to be altered for a monitoring session before connecting
transducers. The changes made using this option do not remain in force when FM800E is
switched off.
See Chapter 9.
4.3
46B
Default alarms
This option allows the default alarms for the FM800E to be altered. The changes made
using this option
do
remain in force when the FM800E is switched off.
See Chapter 9.
4.4
47B
System settings
See Chapter 5.