Sonicaid FM800 Encore Series Instructions for Use
73
9
8B
Setting Alarm Thresholds
9.1
80B
Default alarm thresholds
FM800E comes with factory-set alarm thresholds for all parameters:
Parameter
Low
High
Status
ULT 1, ULT 2, FECG
110bpm
160 bpm
Off
FHR Signal Loss
None
30%
Off
MECG
50bpm
120bpm
On
Maternal Blood Pressure, Systolic
Maternal Blood Pressure, Diastolic
90mmHg
50mmHg
160mmHg
90mmHg
On
On
Maternal Oximetry
94%
None
On
222B
Changing thresholds
If the current thresholds are changed, the changes remain in force until FM800E is switched
off. If the default thresholds are changed, the changes remain in force permanently. The
screens for changing current and default thresholds are the same. The only difference is the
option required to select to access the screen.
To access the current alarm thresholds screen:
> SETUP
> CURRENT ALARMS
To restore the default thresholds, switch FM800E off and on again.
To access the default alarm thresholds screen:
> SETUP
> DEFAULT ALARMS
> access code (2755)
Changing FHR thresholds is explained in some detail. The procedure for changing other
thresholds is so similar that it is explained much more briefly.