Huntleigh IRT10 Instructions For Use Manual Download Page 7

8

3.  Description of Controls, Indicators,    

 

Symbols and Displays

3.5    Thermometer Controls and Display 

1

2

3

4

5

6

1

Temperature probe status indicator

2

Start measurment

3

Display

4

Infra red sensor

5

Lens ejector

6

Battery compartment

Intr

oduction

Summary of Contents for IRT10

Page 1: ...Wireless Tympanic Thermometer IRT10...

Page 2: ...ts 11 4 3 Connecting the Thermometer to the host 11 4 4 Installing the Batteries 12 4 5 Pairing the Thermometer and Main Unit 12 5 Using the Thermometer 13 5 1 Temperature Measurement 13 6 Care Cleani...

Page 3: ...on History The documentation part number and revision number indicate its current edition The revision number changes when a new edition is printed in accordance with the revision history of the docum...

Page 4: ...use and ensure its safe and proper use WARNING Before use carefully read directions for use including all warnings cautions and instructions WARNING If the battery shows any signs of damage leakage or...

Page 5: ...ital Signs Monitor 2 1 Features of the IRT10 Wireless Themometer Physical The IRT10 wireless tympanic thermometer is a clinical grade hand held thermometer indicated for the intermittent measurment of...

Page 6: ...uld not be used if the following situations are present Do not use the thermometer if there is any blood or drainage in the ear canal Do not use the thermometer if the patient presents with inflamator...

Page 7: ...trols Indicators Symbols and Displays 3 5 Thermometer Controls and Display 1 2 3 4 5 6 1 Temperature probe status indicator 2 Start measurment 3 Display 4 Infra red sensor 5 Lens ejector 6 Battery com...

Page 8: ...ption of Symbols Indicators The symbols and Indicators are described as follows Symbols Description Attention consult accompanying documents Battery charging indicator lamp Refer to instructions for u...

Page 9: ...ormit Europ enne Complies with medical device directive 93 42 EEC Disposal instructions Manufacturer This way Up Fragile Handle with care Limit of stacking layers Keep dry Description of Controls Indi...

Page 10: ...ing and Inspection The Thermometer is shipped in one carton Examine the carton carefully for evidence of damage Contact Huntleigh Healthcare Ltd Service Department immediately if any damage is discove...

Page 11: ...y compartment Place new batteries into compartment paying attention to the polarity close the compartment ensuring it snaps into place The system is ready for use 4 5 Pairing the Thermometer and Main...

Page 12: ...ens cover in accordance with the local regulations During the monitoring process the temperature measuring instrument will automatically check itself once per hour Self checking will last 2 seconds an...

Page 13: ...hat the measurement is complete 5 Remove the thermometer 6 The measurement will be shown on the thermometer s screen it will be automatically transferred to the monitor Thermometer display Monitor dis...

Page 14: ...n is not available use a suitable carton with appropriate packing material to protect the device during shipping Return by any shipping method that provides proof of delivery 6 3 Service Apart from a...

Page 15: ...f range 10 40 C 33 8 39 2 F Move the device to appropriate environment and wait 30 minutes prior to taking a temperature Er2 The environment temperature is not stable Do not take measurement until the...

Page 16: ...The Smartsigns Compact 300 should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary the Smartsigns Compact 300 should be observed to verify norm...

Page 17: ...may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people a The ISM industrial scientific and medical bands between 150 k...

Page 18: ...e s 2 kV line s to earth Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 6100...

Page 19: ...power of transmitter W Separation distance according to frequency of transmitter m 150kHz to 80MHz d 1 2 P 80MHz to 800MHz d 1 2 P 800MHz to 2 5GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1...

Page 20: ...AMMABLE ANAESTHETIC MIXTURE WITH AIR OXYGEN OR NITROUS OXIDE 9 2 Standards MDD 93 42 EEC EN ISO13485 2012 AC2012 EN ISO14971 2012 EN 60601 1 2006 AC 2013 EN 60601 1 2 2007 AC 2010 EN60601 1 6 2010 EN...

Page 21: ...700 hPa to 1060 hPa Note The system may not meet its performance specifications if stored or used outside the manufacturer s specified temperature and humidity range 9 5 Measurement Parameters Probe T...

Page 22: ...aned to the outside of the package Mark the package Service Department For further details refer to NHS document HSG 93 26 UK only Huntleigh Diagnostics reserve the right to return product that does n...

Page 23: ...rjoHuntleigh House Houghton Hall Business Park Houghton Regis Bedfordshire LU5 5XF Registered No 942245 England Wales Registered Office Huntleigh Healthcare Limited 2017 As our policy is one of contin...