Page 1: ...Hydroven 12 LymphAssist Homecare...

Page 2: ...Homecare System 17 7 1 Garment Connections 17 7 2 Operating Modes 18 7 2 1 Standby 18 7 2 2 Sleep Mode 18 7 2 3 Run Mode 18 7 3 Inflation Modes 19 7 3 1 LymphAssist Inflation 19 7 3 2 LymphAssist Syst...

Page 3: ...1 3 Service Period 28 11 2 General Care Maintenance and Inspection 28 11 2 1 Pump 28 11 2 2 Garments 28 11 3 Serial Number Labels 28 11 3 1 Pump 28 11 3 2 Garments 28 12 Trouble Shooting 29 13 Access...

Page 4: ...stem Failure to observe this caution could result in injury or in extreme cases death WARNING A possible explosion hazard exists if used in the presence of flammable anaesthetics WARNING Do not mount...

Page 5: ...nly the pump and garment insert combination as indicated by Huntleigh should be used The correct function of the product cannot be guaranteed if incorrect pump and garment combinations are used WARNIN...

Page 6: ...nt outlet so that the devices are on different branch circuits WARNING The Hydroven 12 should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary t...

Page 7: ...kV line s to line s 2 kV line s to earth 1 kV line s to line s 2 kV line s to earth Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions...

Page 8: ...rs as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of the equipment marked with the followin...

Page 9: ...ransmitter W Separation distance according to frequency of transmitter m 150kHz to 80MHz d 1 2 P 80MHz to 800MHz d 1 2 P 800MHz to 2 5GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 1...

Page 10: ...are detailed in the Indications section 3 3 About the LymphAssist system The system comprises a pump and multi chamber arm or leg garments Optional multi chamber garment inserts can be used to increas...

Page 11: ...n any environment where it may come into contact with water 3 5 Intended User Profile The Hydroven 12 LymphAssist HC is intended for use by all adults providing the clinical condition s identified in...

Page 12: ...mme Lymphoedema Primary and secondary including post surgery radio or chemotherapy Selection should be based upon a holistic assessment of the patients individual care needs Note These systems represe...

Page 13: ...e conditions please consult a physician before use CAUTION IPC should be used with care in patients with the following symptoms or conditions Peripheral neuropathy pain or numbness in the limb Undiagn...

Page 14: ...eceived Should any damage be evident or any parts missing ensure that Huntleigh Healthcare Ltd is informed at once Storage If the unit not be required for immediate use it should be re sealed into its...

Page 15: ...inserts have multiple chambers Each chamber overlaps the adjacent chamber to provide smooth and seamless pressure application and prevent pressure voids or ridging to the limbs All garments have zip...

Page 16: ...Note Garments should not feel tight fitting when deflated 1 Arm Garments The length of the garment is measured from the fingertips to 5 cm below the armpit with the arm straight The circumference of t...

Page 17: ...ions Standard Large Garment Part No 316L76S 316L84S 316L76W 316L84W Length top of thigh to heel 76 cm 84 cm 76 cm 84 cm Circumference at top of thigh Leg Garment with Garment Insert 71 cm 90 cm 71 cm...

Page 18: ...ed to these two 12 way connectors which are push fit and polarised to prevent incorrect orientation of the connectors The following pump garment connections are possible If two garments are used eithe...

Page 19: ...Standby for 10 minutes without any of the buttons being pressed it then goes into Sleep mode to conserve power Note The pump can be left in Sleep mode for long periods without disconnecting the mains...

Page 20: ...of the garment between the next chamber and chamber 12 For example at chamber 5 the 5 LymphAssist pulses are followed by a partial peristaltic Wave inflation from chamber 6 to chamber 12 1 2 3 4 5 6...

Page 21: ...sure that can be set in LymphAssist mode is 40 mmHg 7 3 2 LymphAssist System The LymphAssist system is factory set to deliver two LymphAssist cycles lasting approximately 35 minutes The inflation time...

Page 22: ...tting buttons Pressure Display Run mode indicator Sleep mode indicator Run Standby button Therapy time display 8 2 Pump Controls and Indicators The pump control panel has the following controls and in...

Page 23: ...button after the pump has stopped to reset it to Standby If the pump has been in Standby for 10 minutes with no button being pressed it then goes into Sleep mode to conserve power Only the Sleep mode...

Page 24: ...he end of the therapy as set by the Treatment duration the pump will stop 8 3 Pump Alarm 8 3 1 System Fault 1 If a system fault is detected the pump will stop operating 2 An audible warning will sound...

Page 25: ...1 Remove the system from the packaging You should have the following items Hydroven 12 LymphAssist pump Hydroven 12 garment arm or leg Hydroven 12 garment insert optional 2 Make sure the garment is th...

Page 26: ...o the lower connector Note The connectors are push fit and polarised to prevent incorrect orientation 8 If the patient is setting up or operating the system on his her own make sure that The pump is s...

Page 27: ...b unless it is partially zipped as you may damage the garment zip CAUTION Do not undo the zips or attempt to remove the garments during the therapy session or you may damage the zips Make sure the the...

Page 28: ...not spray cleaning solutions directly onto the pump Do not immerse the tubeset in water 10 1 Cleaning Clean all exposed surfaces and remove any organic debris by wiping with a cloth moistened with a...

Page 29: ...nt 11 2 General Care Maintenance and Inspection 11 2 1 Pump Check all electrical connections and the mains power cord for signs of excessive wear In the event of the pump being subjected to abnormal t...

Page 30: ...ngineer Pump operates but garment will not inflate Blockage in garment supply tube Garment not fitted correctly to pump Air leak in garment Ensure that the tube airway is clear Check connections Check...

Page 31: ...ons Standard Large Garment Part No 316L76S 316L84S 316L76W 316L84W Length top of thigh to heel 76 cm 29 9 84 cm 33 1 76 cm 29 9 84 cm 33 1 Circumference at top of thigh Leg garment 71 cm 28 0 71 cm 28...

Page 32: ...5mm diameter and water droplets falling vertically Degree of safety of application in the presence of a flammable anaesthetic Equipment not suitable for use in the presence of a FLAMMABLE ANAESTHETIC...

Page 33: ...he pump must be allowed to adjust to normal temperatures for a minimum of 12 hours before use Failure to do so may result in accelerated wear of mechanical components 14 4 Standards Compliance EN60601...

Page 34: ...2 2 No 60601 1 2008 MEDICAL EQUIPMENT This symbol signifies that this product including its accessories and consumables is subject to the WEEE Waste Electrical and Electronic Equipment regulations and...

Page 35: ...ing Cleaning Symbols Wipe surface with damp cloth C 1000ppm NaOCl NaDCC Use solution diluted to 1000 ppm of Available Chlorine Do not iron Do Not Use Phenol based cleaning Solutions Do not dry clean D...

Page 36: ...isposal This symbol signifies that this product including its accessories and consumables is subject to the WEEE Waste Electrical and Electronic Equipment regulations and should be disposed of respons...

Page 37: ...contamination certificate or other statement declaring that the product has been cleaned to the outside of the package Mark the package Service Department For further details refer to NHS document HSG...

Page 38: ......

Page 39: ...h A S Vassinger dvej 52 DK 3540 LYNGE Tel 45 49 13 84 86 Fax 45 49 13 84 87 E mail dk_kundeservice arjohuntleigh com USA ArjoHuntleigh Inc 2349 W Lake Street Suite 250 US Addison IL 60101 Tel 1 630 30...

Page 40: ...us improvement we reserve the right to modify designs without prior notice and are trademarks of Huntleigh Technology Limited A Member of the Arjo Family AW 1001008 4 As a proud member of the Arjo fam...

Reviews: