36 Replacing the Pressure Control Assembly
Note : It is
not
necessary to remove the side panel assembly from the pressure control module.
36.1 Remove the pressure control module from the pump case as described in Section 29.
36.2 Remove the pressure control flaps as described in Section 35.
36.3 Release the control adjustment screw on the control knob arbour (Fig 20, Item 10).
36.4 Carefully remove the arbour, cam, spring cup and spring (Fig 20, Items 20, 30 and 40).
36.5 Renew/replace components with the same type.
Note : Early versions may have a nylon washer in lieu of the spring cup. Discard the washer and
replace with the cup.
36.6 Replacement is the reverse of the above procedure. Refit the pressure control flaps as
described in Section 35.
Note : The new control knob arbour will NOT fit the old pressure assembly module because of the
moulded wall on the old pressure assembly module. The old control knob abour will fit both.
Issue 2
Chap 3
Nov 2001
SER0004
Page 39
Mini Pump Service Manual
Pump Repair
Summary of Contents for AlphaBed
Page 1: ...Issue 2 SER0004 Nov 2001 Mini Pump Service Manual Huntleigh HEALTHCARE...
Page 2: ......
Page 4: ...Contents Issue 2 Page ii SER0004 Nov 2001 Contents Mini Pump Service Manual...
Page 8: ...Contents Issue 2 Page iv SER0004 Nov 2001 Contents Mini Pump Service Manual...
Page 16: ...Chap 1 Issue 2 Page 8 SER0004 Nov 2001 Introduction Mini Pump Service Manual...
Page 20: ...Chap 2 Issue 2 Page 4 SER0004 Nov 2001 Maintenance Mini Pump Service Manual...
Page 82: ...Chap 5 Issue 2 Page 8 SER0004 Nov 2001 Testing Mini Pump Service Manual...
Page 86: ...Chap 6 Issue 2 Page 4 SER0004 Nov 2001 Technical specification Mini Pump Service Manual...
Page 92: ...Chap 7 Issue 2 Page 6 SER0004 Nov 2001 Parts list Mini Pump Service Manual...
Page 94: ...Chap 8 Issue 2 Page 2 SER0004 Nov 2001 Service Contact Details Mini Pump Service Manual...