EasyMatch QC User’s Manual Version 2.1
The Sensor Menu
8-7
Sensor/Set Interval
The
Set Interval
command in the
Sensor
menu allows you to set the standardization interval for the
currently active sensor. When that interval has elapsed, the software will prompt you to standardize
before more readings can be made. The interval indicated here applies to EasyMatch QC as a whole,
regardless of the job currently in use.
Note: A standardization interval of no longer than 8 hours is recommended.
If your installed sensor is a LabScan XE, you may also set the secondary standardization interval, which
is the interval at which the secondary white tile “paddle” is automatically checked. Zero disables
reading of this tile entirely.