Humboldt HS-4157 Product Manual Download Page 4

3

2.3. OFF XX (Automatic Shutdown) 
This setting is used to set the time for automatic shutdown. An option of 

between "0-30" minutes can be selected. If "0" minutes is selected, the 

device will not turn off automatically.
2.4. CAL XX (Calibration) 
This setting is used to calibrate. There are two types of calibration in this 

device. One for the bar unit and the other for the other units (kg, tone, 

kN and N) 
2.4.1. Calibration for bar unit 
When calibrating the bar unit, first select "CAL 05" to set the full scale 

value. This setting is 350 bar for the bar unit. Then "CAL 06" should be 

selected for calibration of the device. The CAL 06 option has calibration 

points in it. Here, the device will ask for 6 points between zero and full 

scale 
(including 0). 
First the display will show "0 000" calibration point. The system should 

not have loads at this point. Press the "zero" button and continue. After 

this zero point the device will show 20% of the full scale. In this step, 

a reliable load cell must be applied to the device by measuring the 

pressure load cell that the user requests. When the correct pressure is 

reached, press the "zero" button to continue. This procedure will be 

repeated 6 times in total. The pressure points are as follows:

“1 0XX”… %20 “2 0XX”… %40 “3 0XX”… %60 “4 0XX”… %80

“5 0XX”… %100 (full scale) 
After pressing the zero button on the "5 XXX" digit, the instrument will 

display the calibration coefficient and the calibration will end. 
2.4.2. Calibration for other unit 
For other units, the calibration process is similar to the calibration 

process for the "bar" unit. For this calibration, the unit that is required 

first is to be selected. Once the unit has been selected, the 
"CAL15" option must be selected to set the exact scale. Then "CAL16" 

should be selected. The CAL16 menu contains calibration points. You 

will see a total of 6 calibration points just like on the CAL06 menu. After 

this point, all operations are the same as the bar calibration. After the 

calibration is finished, set "CAL 00" to zero of the device and exit from 

Summary of Contents for HS-4157

Page 1: ...product manual 06 18 HS 4157 Point Load Tester ...

Page 2: ......

Page 3: ...ds 2 1 Uni X Units The user can view the instrument s measurement with 5 different units The SET button must be pressed to change the unit After viewing Uni X on the screen it can be easily changed with the up arrow button All units are given below together with unit numbers Uni 0 bar Uni 1 kg Uni 2 tone Uni 3 Newton Uni 4 kN 2 2 FL XX Filter The filter number is the number of samples that the dev...

Page 4: ...vice will show 20 of the full scale In this step a reliable load cell must be applied to the device by measuring the pressure load cell that the user requests When the correct pressure is reached press the zero button to continue This procedure will be repeated 6 times in total The pressure points are as follows 1 0XX 20 2 0XX 40 3 0XX 60 4 0XX 80 5 0XX 100 full scale After pressing the zero butto...

Page 5: ... This is some kind of fine tuning Unloaded state No pressure load on the device PEAK When this button is pressed the device will show the maximum pressure point at that time For example Assume that we make a pressure measurement as in Figure 1 and press the peak button at different times First time interval A B The measurement results are increasing in the first time interval So whichever point in...

Page 6: ...5 ON SET This button is used to switch the device ON and OFF and also it is used to change the menu ...

Page 7: ...7 ...

Page 8: ...Humboldt Mfg Co 875 Tollgate Road Elgin Illinois 60123 U S A U S A Toll Free 1 800 544 7220 Voice 1 708 456 6300 Fax 1 708 456 0137 Email hmc humboldtmfg com ...

Reviews: