3
2.3. OFF XX (Automatic Shutdown)
This setting is used to set the time for automatic shutdown. An option of
between "0-30" minutes can be selected. If "0" minutes is selected, the
device will not turn off automatically.
2.4. CAL XX (Calibration)
This setting is used to calibrate. There are two types of calibration in this
device. One for the bar unit and the other for the other units (kg, tone,
kN and N)
2.4.1. Calibration for bar unit
When calibrating the bar unit, first select "CAL 05" to set the full scale
value. This setting is 350 bar for the bar unit. Then "CAL 06" should be
selected for calibration of the device. The CAL 06 option has calibration
points in it. Here, the device will ask for 6 points between zero and full
scale
(including 0).
First the display will show "0 000" calibration point. The system should
not have loads at this point. Press the "zero" button and continue. After
this zero point the device will show 20% of the full scale. In this step,
a reliable load cell must be applied to the device by measuring the
pressure load cell that the user requests. When the correct pressure is
reached, press the "zero" button to continue. This procedure will be
repeated 6 times in total. The pressure points are as follows:
“1 0XX”… %20 “2 0XX”… %40 “3 0XX”… %60 “4 0XX”… %80
“5 0XX”… %100 (full scale)
After pressing the zero button on the "5 XXX" digit, the instrument will
display the calibration coefficient and the calibration will end.
2.4.2. Calibration for other unit
For other units, the calibration process is similar to the calibration
process for the "bar" unit. For this calibration, the unit that is required
first is to be selected. Once the unit has been selected, the
"CAL15" option must be selected to set the exact scale. Then "CAL16"
should be selected. The CAL16 menu contains calibration points. You
will see a total of 6 calibration points just like on the CAL06 menu. After
this point, all operations are the same as the bar calibration. After the
calibration is finished, set "CAL 00" to zero of the device and exit from