ENG
LIS
H
www.humancaregroup.com
17
Troubleshooting
Problem
Solution
The lift will not operate the up/down function.
Check that the emergency stop is off.
Check that the battery is sufficiently charged.
The lift will not operate the open/close function.
Check that the emergency stop is off.
Check that the battery is sufficiently charged.
Ensure nothing is blocking the path.
The hand control will not work.
Check that the plug is properly inserted into the
control box.
Check the control box to see if the lift functions.
If it is properly functioning, replace the hand
control.
The lift will not start.
Check that the plugs are properly inserted into
the control box and mains.
The battery is not charging.
Release the emergency stop button.
Maintenance
The Levanto is constructed in such a way that maintenance can be kept to a minimum. It is recom-
mended that the following maintenance scheduled is maintained.
Checklist Prior to Each Use
ü
Battery is fully charged
ü
Sling shows no signs of wear
ü
Plastic components i.e. clips are not
cracked, damaged or worn
ü
Electrical lift and hand control functions
checked and working
ü
Brakes hold position when locked
Checklist After Each Use
ü
Clean according to cleaning instructions.
Weekly Maintenance
ü
Check the operation of the lift and the
operation of the emergency stop.
ü
Check that all parts move easily and
without any noise.
ü
Charge the batteries.
Semi-Annual Maintenance
ü
Complete the service checklist. If the lift
does not meet one or more of the checklist
points, it should be temporarily deactivated
from use until such time that it meets the
maintenance criteria.
Annual Maintenance
ü
Once a year, according to ISO 10535, the
lift must be inspected. This inspection must
be done only by qualified and authorized
personnel. Each lift shall have a service control
sticker affixed indicating the date of inspection.