producing technology for health
13-
Maintenance and
reconditioning
The bed is designed to require the minimum maintenance. It is
recommended that preventive maintenance be carried out according to our
service schedule every other year.
Normal Cleaning
Mechanics and motor systems are cleaned with a damp cloth. Readily
available cleansers and disinfectants can be used. Can be rinsed with
spray water.
The bed ends and long sides are cleaned with a damp cloth. Readily
available cleansers and disinfectants can be used.
High-pressure Wash
Before washing motors have to be in the bottom position. Lift out the
electronics unit.
Inspection in the Case of Reconditioning
Check that the attachment points for all motors are undamaged and do
not have unnaturally large gaps.
Check that there isn't any visible damage on the motors and mechanics.
Check that joints and fasteners for the back and leg sections are
undamaged and do not have unnaturally large gaps.
Check that joints and fasteners for the bed lift's scissor section are
undamaged and do not have an unnaturally large gap.
Check that the bed lift's wheels are firmly tightened.
Tighten the screws for bed ends and long sides.
Check that the mains cable and cable to the control unit are undamaged.
Re-lubricate if necessary according to inspection report.