6  Integration into Automated Test Systems

For integration into automated test systems the commands of the USB interface are described in 
the following. The USB interface is installed as a virtual COM port at 9600 baud, 8 data bits, 1 stop
bit and no parity. The command frame consist of a length byte (entire frame) followed by an 
address byte. Then there is a command byte and optional parameters. Only command frames 
corresponding to its own address will be considered. There is a timeout of 500 ms, that is, if after 
500 ms the required number of bytes hasn't been sent, the previous bytes are ignored! The 
feedback frame corresponds to the command frame, e.g. length, address, command and optional 
parameters. The feedback is always carried out 

after

 execution of the command! When an 

unknown command occurs, the feedback is 0xFE, and with a timeout 0xFD. Several amplifiers can
be driven via a COM interface. In this case, each amplifier requires its own address (1 ... 99, default
= 1), which is determined by a command. The address 100 is answered by all amplifiers regardless
of the stored address. 

6.1  List of Remote Commands

0x01

Status query, parameters: no; answer: 1 Byte status:
Bit 0: Ready
Bit 1: Overload shutdown power/power loss

  

Bit 2: Heatsink overtemperature/transformer
Bit 3: Output relay (0: off, 1: on)
Bit 4: 50R-Input relay (0: off, 1: on)
Bit 5: 100k-Input relay (0: off, 1: on)
Bit 6: operating v (0: low, 1: high)
Bit 7: operating voltage – (0: low, 1: high)

0x02

Set 50R-Input , Parameter: 1 Byte (0: off, 1: on)

0x03

Set 100k-input , Parameter: 1 Byte (0: off, 1: on)

0x04

Set Output relay , Parameter: 1 Byte (0: off, 1: on)

0x05

Set operating voltage, Parameter: 1 Byte

(0:  low, 1: high, 2: only UB+ high, 3: only UB– high)

0x06

Query temperature, Parameter: no, answer: 1 Byte temperature in 

0

C

0x07

Query max. power loss, Parameter: no, answer: 1 Byte 
Power loss in % of the current threshold (maximum since last  
query)

0x08

Query average power loss, Parameter: no, answer: 1 Byte 
power loss in % of the  current threshold

0x09

Query Error, Parameter: no answer 1 Byte Error:

Bit 0: short circuit-current
Bit 1: Overc

A1230-01 

Integration into Automated Test Systems

Page 19

Summary of Contents for A1230-01

Page 1: ...Operating Manual A1230 01 4 Quadrant Voltage Amplifier 2 Channel DC 7 MHz 450 V s 185 W DC source 68 W DC sink Revision 2 0 MA 12300010...

Page 2: ...lements 8 4 A1230 Control Software 11 5 Notes on Operation 16 6 Integration into Automated Test Systems 19 7 Maintenance and Service 22 8 Decommissioning and Disposal 23 9 Warranty 23 10 Contact Infor...

Page 3: ...e device is equipped with a temperature controlled low noise fan Additionally an over temperature protection a power loss calculation and an absolute current monitoring guarantee perfect short circuit...

Page 4: ...or serious injury Type and source of danger Indicates a potentially hazardous situation which if not avoided could result in minor or moderate injury or property damage Type and source of danger Indic...

Page 5: ...k of electric shock the device housing must be earthed and therefore always operate the device via the supplied three core power cable with protective conductor The mains cable may only be plugged int...

Page 6: ...to be performed contact the manufacturer of the device 2 3 Intended Use 2 3 1 Hardware The device is intended exclusively for use as an LF power amplifier Typical applications are the operation as vol...

Page 7: ...ser Operation may only be performed by qualified persons Reading and understanding the operating instructions Never use the device without having read and understood the operating instructions Always...

Page 8: ...itched on 3 AMP IN 100 K BUTTON OFF 100 k input is switched off ON 100 k input is switched on 4 OUT OUT 4 mm safety sockets amplifier output Reference potential negative socket is connected to protect...

Page 9: ...r is ready for operation READY Red Signals the intervention of a protection mechanism Red LED lights up constantly Over temperature protection the device switches on automatically after the drop of th...

Page 10: ...4AT 5x20 mm 2 BRIDGE OUT BNC SOCKET Inverted preamp output for simple setup of bridge circuits 3 INTERLOCK BNC socket for interlock connection 4 COMPUTER USB USB socket for connecting to a computer em...

Page 11: ...tup exe Additional drivers and required additional programs are installed within the framework of the installation routine and require administrative access rights The driver devices required for the...

Page 12: ...artup configuration The amplifier then starts after the repeated switching on with the adjusted options 2 DEVICE saadfasdfasdfsd 3 INFO Here you find information about your GUI software version as wel...

Page 13: ...ction must be reset manually The amplifier can now be switched on again Amplifier On 9 STATUS INDICATION Green Ready Signals the operational readiness of the amplifier Red short circuit_current exceed...

Page 14: ...signal output of the amplifier can be turned on and off by clicking the appropriate box 4 3 2 Connect Menu The available amplifiers are listed with their corresponding port A connection is establishe...

Page 15: ...n of the state of the 100 k input 4 INTERLOCK Setting the interlock mode LATCHING After an interlock the interlock state must be reset by a remote command before the amplifier can be switched through...

Page 16: ...ep the power loss of the amplifier low the operating voltage should always be selected according to the load Example The load current IL should be 2 ADC Load 1 RL 25 For IL 2 ADC an output voltage UA...

Page 17: ...e operation the cables should be as short as possible and of similar length The cable cross section should not fall below 1 5 mm The A1230 01 is a fast broadband amplifier that can deliver fast pulses...

Page 18: ...t amplifier is connected to the 100 k input of the second amplifier Switch the 100 k input of the second amplifier on so you can control the entire output signal via the button of the selected input o...

Page 19: ...1 99 default 1 which is determined by a command The address 100 is answered by all amplifiers regardless of the stored address 6 1 List of Remote Commands 0x01 Status query parameters no answer 1 Byt...

Page 20: ...sion Parameters no answer 1 Byte revision 0x16 Set hardware revision Parameters 1 Byte revision e g 0x21 2 1 0x17 Query hardware revision Parameters no answer 1 Byte revision 0x53 Set Interlock Mode P...

Page 21: ...turer s interface listed below Current devices https www lantronix com products deviceinstaller Older devices https www eztcp com en download ezmanager php Drivers and product information on the curre...

Page 22: ...y Shipping without a pallet usually leads to considerable transport damage the costs of which we do not cover Please include the RMA number with the delivery e g on the delivery note Ensure that the p...

Page 23: ...e product which was not damaged by negligence misuse improper installation accident or unauthorized repair or modification by the buyer This warranty is applicable only to defects due to material or w...

Page 24: ...Bochum Germany Tel 49 234 970569 0 Fax 49 234 970569 29 service drhubert com www drhubert com 11 Guidelines This device complies with the relevant European Union harmonisation legislation EMC Directi...

Page 25: ...12 Document History Revision Date Changelog 1 1 February 2020 First publication in new design 2 0 March 2020 Backpanel and frontpanel images were updated A1230 01 Document History Page 25...

Reviews: