background image

4

alcohol and 500ml petroleum ether into the A container;

Take 1 bottle of extract medicine block (see above) into reaction cup B, add 10ml

isopropanol and 10ml petroleum ether, stir the extract medicine block, isopropanol and

petroleum ether in the reaction cup, stir and dissolve into the pre In the prepared A

container;

Pour 10ml of petroleum ether and 10ml of isopropanol into the B reaction cup. After

mixing petroleum ether, isopropanol and the residue, pour it into the A container again and

stir evenly;

Finally, the mixed liquid in container A is poured into empty glass bottles (empty bottles

containing isopropyl alcohol and petroleum ether can be used), the cap is screwed tightly

and shaken sufficiently to completely dissolve, and the extraction liquid is completed.

3

Extraction

The essential materials include a container(a clean and dry drinking water

bottle is allowed) that more than 1000ml capacity, a reaction cup,500ml Isopropyl Alcohol

and 500ml Petroleum Ether. Open the plastic bag with the extract provided with the

instrument, put the extract in the reaction cup, and add a small amount of isopropyl

alcohol and petroleum ether to dissolve, transfer the solution to the large container

prepared in advance, then wash the reaction cup with petroleum ether and isopropyl

alcohol several times. The cleaning liquid is poured into the large container together. The

remaining petroleum ether and isopropyl alcohol are all poured into the large container

and shaken. At last, inject the solution in the blank bottles which contained Isopropyl

Alcohol and Petroleum Ether.

4

Standard acid solution

The solution is the 0.15mol/l Potassium hydrogen phthalate

solution

which can access directly in the small glass bottle.

5

BTB aqueous solution

Open the bag with BTB provided with the instrument and

transfers it in the 100ml container as soon as possible. Then inject 100ml Anhydrous

Ethanol and shake it completely. Draw 0.5 ml of this solution with the graduated pipette

and add 50ml distilled water, and then there is the BTB aqueous solution.

6

Washing solution

Transfer the white solid in the glass bottle labeled with KOH to the

wash gas cylinders as soon as possible. Then inject 50ml water and shake it

completely .At last set aside to use.

Summary of Contents for HZCS-3

Page 1: ...HZCS 3 Automatic Oil Acidity Tester Huazheng Electric Manufacturing Baoding Co Ltd...

Page 2: ...Contents I Overview and Technical Index 2 II Structure and Installation 3 III Preparation Work 3 IV Instrument Operation 5 V Maintenance 21 VI Operation Notice 22 VII Packing List 23...

Page 3: ...k Please read this User Manual before using the tester and follow the instructions to operate and maintain it for prolonging its service life This instrument is characterized by Performing the test au...

Page 4: ...n the cup Once started 1 to 3 samples can be measured which is convenient and high efficiency Because of the preparation of the standard acid for instrument calibration and calibration procedure the c...

Page 5: ...KOH solution Extraction Standard acid solution BTB aqueous solution Washing solution Anhydrous ethanol 1 Neutralizer Ethanol KOH The white solid in the glass bottle labeled KOH supplied with the instr...

Page 6: ...th the instrument put the extract in the reaction cup and add a small amount of isopropyl alcohol and petroleum ether to dissolve transfer the solution to the large container prepared in advance then...

Page 7: ...affixed with a neutral label with a hose Inject the extract from the reagent bottle into the extraction cell Connect the bottle of anhydrous ethanol to the interface affixed with a anhydrous ethanol...

Page 8: ...f the instrument and place it in the sample cup placed in the second cup position After the sample cup is detected the instrument starts to extract the extract If there is no extract it will prompt to...

Page 9: ...7 When there is 20 drops the instrument enters the main interface Click the time shown on the upper right corner the instrument enters the following interface and time can reset...

Page 10: ...t time or the neutralization is being changed this step is necessary In other situations this step can be ignored In the main interface click Instrument Calibration the instrument enters the following...

Page 11: ...in the chapter 3 BTB aqueous solution and put the cup in the sample tray Click the Start button and a blank titration is beginning The instrument enters the following interface When the blank titratio...

Page 12: ...ut the volume of the standard acid solution inject corresponding 0 15mol L standard acid solution with micro injector see details in the chapter 3 Standard acid solution in the cup Click the Start but...

Page 13: ...er the main calibration interface If verification is in progress the instrument enters the following interface Click the Print button to print the result of verification or go back the main calibratio...

Page 14: ...12 If there is no enough extract the instrument warns like the following interface Add enough extract and press OK after the self inspected the instrument enters the following interface...

Page 15: ...ct The specific steps are to put a clean sample cup into the sample tray Click the Blank button to start blank titration of the extract The instrument enters the following interface Looking for the sa...

Page 16: ...14 Replace the sample cup with a clean sample cup Click OK to start extracting 20ml of extract Then blank titration of the extract is in progress the interface is as following...

Page 17: ...ract should be blank calibrated in the following cases 1 replace the extract 2 Replace the neutralizing solution Then the instrument enters the following interface Input the oil density click the Turn...

Page 18: ...t cup when the first cup is founded and the cup is founded if there is no enough extract the instrument enters the following interface After add enough extract press the OK button to continue testing...

Page 19: ...17 When this step finished the extraction is back flowing the instrument enters the following interface When this step finished the neutralization starts the instrument enters the following interface...

Page 20: ...hed the result is shown and the instrument is looking for the next cup the instrument enters the following interface If there is no sample cup the instrument is looking for the next cup and shows as t...

Page 21: ...e click the History Data Query button the instrument enters the following interface Press the PgUp and PgDn button history data can be founded press the Delete button delete data shown on the screen p...

Page 22: ...s no history data the instrument enters the following interface 5 Instrument cleaning Put in a sample cup and press the Instrument Cleaning button in the main interface the instrument starts cleaning...

Page 23: ...and inject it into the cell 3 Configure a bottle of neutralization when it run out and take out the bottle labeled with neutralization inject the neutralization into the bottle labeled with neutraliza...

Page 24: ...o wear Please check whether the hose is damaged or adhesion carefully to reduce the impact on the instrument If there is any abnormal replace the hose After the instrument finished please release the...

Page 25: ...liquid medicine 2 3 Extracted liquid medicine 2 4 Gas washing bottle medicine 1 5 Indicator 1 6 Standard acid 1 7 Microsyringe 1 8 Sample cup 4 9 Stirrer 5 10 Glass bottle 3 11 Glass bottle cap 3 12...