HUA YI ELECTRONICS HYelec HY3002-2 User Manual Download Page 6

                                                                                               -     - 

5

(20)  Output negative terminal of slave circuit. 

(21)  Mains power on/off switch. 

(22)  Fixed 5V positive output terminal. 

(23)  Fixed 5V negative output terminal. 

2.2    USING METHOD 

2.2.1     DUAL POWER SUPPLY USED IN INDEPENDENT MODE 

2.2.1.1  Move switches 13 & 14 to the up position 

2.2.1.2  For constant voltage mode, adjust controls 6&8 clockwise to the maximum position. 

Move switches 2&3 to the voltage display position. Switch on the power ON/OFF switch and 

adjust controls 5 & 7 to the desired voltage for the master and slave outputs. 

2.2.1.3   For constant current mode, switch on the power ON/OFF switch and adjust controls 5&7 

clockwise to the maximum position. Adjust controls 6&8 anti-clockwise to the minimum position. 

After connecting the load, move switches 2&3 to the current display position. Adjust controls 6&8 

to give the desired current at the master or slave outputs. 

2.2.1.4  For current limit protection mode, switch on the power ON/OFF switch and adjust 

controls 6 & 8 anti-clockwise to the minimum position. Adjust controls 5&7 to give the desired 

output voltage and then connect   output terminals 15 with 17 and 18 with 20 respectively. Move 

switches 2&3 to the current display position. Adjust controls 6&8 clockwise to set the desired 

value of current limit for the master or slave outputs. Remove the connection between 15 & 17 

and 18 & 20, and connect the load.

  

Summary of Contents for HYelec HY3002-2

Page 1: ...USER S MANUAL DC REGULATED POWER SUPPLY...

Page 2: ...e allowing twice the voltage or current and outputs can be tracked relative to the master output The highest output voltage is the sum of twice the voltage value in series mode and the highest output...

Page 3: ...LTAGE 2 0 30V 2 0 30V 2 0 30V 2 0 30V 2 0 30V 2 0 30V OUTPUT CURRENT 2 0 2A 2 0 3A 2 0 5A 2 0 2A 2 0 3A 2 0 5A FIXED OUTPUT NO NO NO 5V 3A 5V 3A 5V 3A MODEL HY5002 2 HY5003 2 HY5005 2 HY6002 2 HY6003...

Page 4: ...urrent indication accuracy LED LCD 2 2digits analogue display 2 5 1 3 Fixed output 1 3 1 Output voltage 5V 2 5 1 3 2 Output current 3A 1 3 3 Line regulation CV 0 01 1mv 1 3 4 Load regulation 0 1 1 3 5...

Page 5: ...for master output 11 Constant voltage mode indicator light for slave output 12 Constant current mode indicator light for slave output and twice current output indicator light in parallel mode 13 Indep...

Page 6: ...OFF switch and adjust controls 5 7 clockwise to the maximum position Adjust controls 6 8 anti clockwise to the minimum position After connecting the load move switches 2 3 to the current display posi...

Page 7: ...terminals 17 18 should be connected together with a thick conductive wire to prevent damage to the unit in the event of an overload 2 2 2 4 In series mode remove any connection between the master or...

Page 8: ...ls however the unit should be turned off and the short circuit removed before continuing use 3 2 The mains power must be switched off before servicing and servicing should be referred to a qualified p...

Reviews: