Measurement Procedure 

1.

 

Long press the On/Off button to turn the gauge on. 

2.

 

Check  the  specimen  and  select  a  suitable  calibration  method.  Calibrate 
accordingly and the latest calibration will be stored and applied as a rule. 

3.

 

Set upper or lower limit when necessary. 

4.

 

Set statistics, storage and adjust other settings when necessary. 

5.

 

Start measurement. Note that the probe should be put perpendicular to the 
surface to be measured to minimize measurement error. 

6.

 

Turn off the gauge after measurement by long press the On/Off button. 

7.

 

Calibration   

1. Types of calibration 

1)

 

Zero calibration: single point, use the zero plate directly. 

2)

 

2 point calibration, smooth: suitable for coatings on smooth surface. 

3)

 

2 point calibration, rough, method 1: recommended for sand- or shot-
blasted  surfaces  with  roughness  of  less  than  55-65  µm.  Calibration 
should be done directly on the blasted surface without coating. 

4)

 

2 point calibration, rough, method 2: recommended for sand- or shot-
blasted  surfaces  with  roughness  of  more  than  55-65  µm.  Calibration 
should be done directly on the blasted surface without coating. 

5)

 

Factory calibration, can be reset by restoring factory setting. 

Note: As to the selection of standard film, one that is slightly thicker than the 
coating to be measured should be used for calibration. For example, for a 
coating  of  about  80  µm,  a  standard  film  of 125 µm  should  be applied  for 
calibration. 

Summary of Contents for CO600 Economic

Page 1: ...Operation Manual HTech Messtechnik GmbH Germany www htech mt de info htech mt de...

Page 2: ...List 2 4 Before First Use 2 5 On Screen 3 6 Measurement Procedure 4 7 Calibration 4 8 Upper Lower Limit Setting 5 9 Statistics 6 10 Language Volume Screen Rotation and Restore Factory Setting 6 11 Not...

Page 3: ...g on non magnetic metal substrates Standards ISO 2178 2360 2808 ISO 19840 ASTM B244 B499 D1186 D1400 BS3900 C5 2 Construction Plan 1 Strap hole 10 Probe socket 2 Light sensor 11 Probe 3 LED indicator...

Page 4: ...Before First Use 1 Check the package according to the packing list 2 Install battery release the screw and remove the cover Insert 2 AA batteries and then fix the cover Note that the battery should be...

Page 5: ...ibration method 2 point calibration smooth Zero calibration 2 point calibration rough Factory calibration 6 Upper Lower limit symbol appears when the limit setting was switched on 7 Exceeding Upper Lo...

Page 6: ...calibration single point use the zero plate directly 2 2 point calibration smooth suitable for coatings on smooth surface 3 2 point calibration rough method 1 recommended for sand or shot blasted surf...

Page 7: ...o the gauge 6 When 2 point calibration smooth finished appears on the screen click OK 7 Calibration finished Note It is recommended that the calibration be performed with blasted substrate as support...

Page 8: ...tart new statistics to clear available data and start new statistics 10 Language Volume Screen Rotation and Restore Factory Setting 1 Can all be adjusted using the Menu on main screen 2 Shortcut for l...

Page 9: ...en 2 2 TFT Battery AA size 2X rechargeable ones applicable Battery life time 20 hours typically may vary with operating conditions Operating temperature 10 to 50 oC Precision 3 Measurement speed Up to...

Page 10: ...000 1 000 000 USB to PC Hardware update WIFI 14 Typical Problems Failure Possible reason Measures Cannot turn the gauge on Battery empty or battery inserted in wrong direction Check battery Fluctuati...

Page 11: ...00 S3000 3000 m standard film CO600 S4000 4000 m standard film CO600 S5000 5000 m standard film CO600 S7000 7000 m standard film CO600 S10000 10000 m standard film CO600 S set Standard film set 50 125...

Page 12: ...us probe CO600 NF5 0 5000 m non ferrous probe CO600 NFR5 0 5000 m rectangular non ferrous probe CO600 NFH5 0 5000 m high temperature non ferrous probe CO600 NF14 0 14000 m non ferrous probe 3 Others P...

Page 13: ...HTech Messtechnik GmbH Von Waldthausen Str 28 45739 Oer Erkenschwick Germany Tel 49 2368 9040942 www htech mt de Email info htech mt de...

Reviews: