Digital Insulation Tester 

1. SPECIFICATIONS 

1.1 General Specifications 

Display:                            3½ digit liquid crystal display (LCD) 20mm height with a maximum reading of 

1999. 

Over range Indication: (1) or (-1) is displayed at MSD  

Low Battery:                  Displayed when the battery voltage drops is Low. 

Safety standard:           Meets the requirement of IEC 1010 for installation category II, 1000V

 

phase 

to earth.   

Measurement rate:      2.5 measurements per second. 

Operating condition:  0°C to + 50°C, 0-75% RH 

Storage conditions:   -20°C to + 60°C, 0-80% RH with battery removed. 

Accuracy:                     + (3% rdg + 2dgt ) Accuracy specifications at 23 +  5°C, less than 75% RH. 

Power Supply:              9 volts (6 x 1.5V) battery, Eveready type 1015 or equivalent. OR Power 

Adaptor 9V, 1.5 A 

Dimensions:                   139 x 99 x 54mm (approx) 

Weight:                          395gms including battery, Eveready type 1015 or equivalent. 

Accessories:                  Test leads (pair) X 1, 1.5V battery (installed) X carrying case X 1. 

Dielectric Strength:      3.5kV @ 50Hz for 1 min. between input terminals and case. 

Insulation Resistance:  

More than 50MΩ at 500V between circuit and case

1.2 Electrical Specifications 

Accuracies are +(% reading plus number of digits) at 23 +

 

5°C, and humidity of less than 75% RH

 

 

      

 

      

     

    

 

 

Model NO 

DIT90A 

DIT90B 

DIT90C 

DIT90D 

DIT90E 

 

Rated Voltage 

1000V 

1000V 

500V 

500V 

 

 

Range 

2000MΩ

 

200MΩ

 

2

000MΩ

 

200MΩ

 

 

 

Resolution 

1

MΩ

 

0.1

MΩ

 

1

 

MΩ

 

0.1

 

MΩ

  

 

 

Max. No. load 
Voltage 

+25% of rated voltage 

 

 

Short Circuit 
Current 
Accuracy

 

<2 mA 

+ (3%rdg + 2dgt) 

 

 

    

2. SAFETY INFORMATION                                                                                                                                                                      

1. The circuit under test must be de energized and isolated before connections are made. 

2. Circuit connections must not be touched during a test. 

3. After insulation tests capacitive circuits must be allowed to be discharged. 

4. Test leads must be in good order, clean and have no cracked or broken insulation. 

5. Do not push the test button before all connections and preparations are done. The instrument 

must only be used by suitable trained and competent persons.  

  

Warning and safety symbols  

 Caution refer to this manual before using the meter.   

 Dangerous 

voltages. 

C€ Comply with IEC1010 –

 1.                                                                                

  

3. OPERATION

 

Before taking any measurements, ready the safety information section. Always 
examine the instrument for Range, contamination (excessive dirt, grease, etc.) 
and defects Examine the teat leads for cracks or frayed insulation. If any abnormal 
conditions exist do not make measurements. 

3.1 PRECAUTIONS 

1. Disconnect all power from the circuit under test when making measurements. If any voltage is 

present in the test circuit an erroneous reading will result. 

2. When the test switch is pressed a high voltage is generated between ‘

L

’ and ‘

E

’ terminals. Do not 

  

touch any live parts in the circuit during measurement. 

3.2 INSULATION RESISTANCE MEASUREMENT 

1. Connect red test lead to ‘

L

’ Terminal

 

2. Connect black test lead to ‘

E

‘ Terminal

 

3. Short the red and black test lead and press t

he test switch, Zero MΩ should be indicated on the 

display. 

4. Connect the test leads to circuit under test and press test switch. 

5. Display indicates the reading in MΩ, if insulation resistance is higher than the instrument, ‘1’ or ‘

-

1’ is

 indicated.     

 

4. MAINTENANCE 

                                                                                                                                                                                                        

If there appears to be a malfunction during the operation of the meter. The 
following steps should be performed in order to isolate the cause of the problem. 

1. Check the battery. 