2. PRODUCT DESCRIPTION
The luxmeter probes are illuminance to
voltage converters. The probes are
battery operated and designed to provide
a direct Luxmeter reading when
connected to any high impedance digital
multimeter (DMM) that accepts shrouded
banana plugs (HT53) and Hypertac
connector (HT53/05) and has 1mV DC
resolution and at least 2000mV DC full
scale readout capability.
The Luxmeter probe provides a fast and
convenient way to measure illuminance in
normal ambient conditions. It is ideal for
spot surveying in any application where
illuminance measurements are important.
The output is 1mV DC per 0.01lux at
20lux range and 1mV DC per 1lux at
2000lux range and 1mV DC per 10lux at
20000lux range (selected with a switch).
3. PREPAPATION FOR USE
3.1. SUPPLY
VOLTAGE
The probe is battery supplied; it uses a
single battery model 9V NEDA 1604, IEC
6F22, JIS 006P included in packaging.
The Battery autonomy is 220 hours.
To avoid battery discharging the
instrument is shipped with battery outside
of the probe; to insert the battery please
follow the indicator of paragraph 5.3.
The “Low BATT” LED indicator light up,
when the battery is nearly discharged. In
case replace it following the instructions
in paragraph 5.3.
3.2. CALIBRATION
The instrument fulfils the technical
characteristics listed in this manual. The
performance of the specification is
guaranteed for one year.
Summary of Contents for HT53
Page 1: ...HT53 HT53 05 Manuale d Uso User manual Manual de instrucciones ...
Page 2: ......
Page 3: ...Indice generale General index Índice general ITALIANO IT 1 ENGLISH EN 1 ESPAÑOL ES 1 ...
Page 4: ......
Page 5: ...ITALIANO Manuale d uso ...
Page 15: ...ENGLISH User manual ...
Page 25: ...ESPAÑOL Manual de instrucciones ...
Page 35: ......