Page 19
HUGO SACHS ELEKTRONIK - HARVARD
A P P A R A T U S
GmbH D-79232 March-Hugstetten Germany
HSE PLUGSYS
®
TAM-A Type 705/1 and TAM-D Type 705/2
9.2 Technical data TAM
Construction
PLUGSYS module
Differential input with an input impedance of 10
10
Ohm through a 6-pin Binder or Amphenol-
Tuchel socket with screw lock. For direct operation of resistive transducers with half-bridge
or full-bridge circuit.
Input
Supply
+5 Volt 350 mA from PLUGSYS system bus
CAL reference
In position CAL, simulate a reference value adjustable by multi-turn trimmer, additionally
zero reference in position OFF
Binder Type 09-0321-00-06 / Amphenol Type T3400501
Input plug
Bridge
supply
5 Volt / 50 mA max.
In the range 10x to 10 000x, selected by internal jumper as x10, x20, x50, x100, x200,
x500, x1000, x2000, x5000 and x10000. The basic gain setting can be varied with the
GAIN control within the range 20-100%.
Gain
1. Autozero by pushbutton
2. Manual zero through multi-turn coarse and fine trimmers
Zeroing
Analogue LED bargraph, 19 LEDs (+12/null/-6) for visual indication of the measurement
signal
Indication TAM-A
3½-digit LED display ±10 Volt / ±100%, or variable to suit the actual measurement
Indication TAM-D
Low-pass with 3 settings 30 - 100 - 300 Hz selected by switch
Output filter
Internally to PLUGSYS system bus: pulsatile output and additional mean value output (0.1
or 0.3 Hz)
Outputs
BNC socket on front panel: pulsatile output or mean value output (0.1 or 0.3 Hz) as se-
lected internally through jumper
19" PLUGSYS module, width 8E (40.5 mm), height 3U (128.7 mm), depth Eurocard (220
mm)
Dimensions
DIN 41612, 96 pin VG
Connector
260g
Weight
Operating Instructions
Accessory