HP HP 8901A Product Overview Download Page 6

6

Special Functions

The modulation analyzer can do
more than is apparent from the
front panel. This capability is
accessed by using the data keys
and a special function key or
remotely via HP-IB. They fall into
three categories: manual control
of instrument functions, instru-
ment operation verification, and
service aids.

An example of the type of special
function found in the manual
control category is the auto
tune-track mode. This mode is
accessed by entering 4.1, then
pressing the special key. Once
the modulation analyzer has been
placed in track mode the analyzer
will continuously track the signal
as it changes frequency. This
eliminates the delays caused by
the analyzer searching for the
signal each time the signal’s
frequency changes. Using this
special function, a user could

continuously monitor modulation
accuracy on a signal generator
while tuning across the signal
generator’s frequency band.
Auto tune-track makes it possible
to tune the analyzer three ways:
automatically, manually by
entering the frequency on the
keyboard, and track mode.

Special functions can also be used
to set any measurement range or
instrument function. They can be
used to select either of two inter-
nal IF’s, the one normally used for
frequencies above 10 MHz or a
narrow IF where rates and devia-
tions are more restricted but
selectivity is increased. All instru-
ment functions not set using these
special functions remain in auto-
matic mode. This allows the user
to select any combination of
manual or automatic operation.
By depressing the special key, the
display shows an eight-digit num-
ber that indicates which functions
are in automatic and the state of
those manually set.

There are also numerous special
functions that can be used in veri-
fying that the instrument and its
various sections are operating
properly. These, along with ser-
vice functions used in diagnosing
and repairing the analyzer, make
repairs much faster and easier.
An additional service aid is the
built-in ability to use HP signature
analysis instrumentation. This
allows a technician with little
knowledge of digital circuits to
rapidly troubleshoot a failure in the
digital portion of the instrument.

Those special functions that are
most commonly used in operating
the analyzer are described on the
pullout information card under the
front panel.

Summary of Contents for HP 8901A

Page 1: ...distortion Second the optional calibrators provide a precise modulation source for calibration There are many metrology and maintenance uses for the modula tion analyzer They include Signal generator calibration Modulation calibration standard VCO and VCXO characterization Applications Frequency Range 150 kHz to 1300 MHz Resolution 10 Hz below 1000 MHz 100 Hz above 1000 MHz Input Level Automatic M...

Page 2: ...tic system to make all these measurements under remote control Single Key Measurements The HP 8901A Modulation Analyzer features easy to use controls In automatic operation all major func tions are selected by pushing a single key No manual tuning or range selection is needed The modu lation analyzer automatically tunes to the input signal adjusts for proper signal level selects the appropriate me...

Page 3: ... 1000 MHz Sensi tivity is 25 dBm 12 mVrms below 650 MHz and 20 dBm 22 mVrms above 650 MHz Besides normal frequency measurement capabilities the analyzer s counter has several additional capabilities Like most frequency counters it will mea sure signals over a wide dynamic range 50 dB 22 mVrms to 7 Vrms and is protected from damage for signals up to 35 Vrms Unlike many frequency counters however it...

Page 4: ...tec tion to prevent damage from the accidental application of exces sive power This is a common cause of damage in equipment such as the modulation analyzer used to measure transmitters The modulation analyzer is tested for inputs up to 25 watts Protec tion is provided by limiting diodes and an RF relay When excessive power is applied the relay opens and protects sensitive compo nents and the anal...

Page 5: ...o be accurately measured The modulation analyzer contains four de emphasis networks that can be used in addition to the audio filters These are the ones commonly used in FM communi cations 25 50 75 and 750 µs When selected the de emphasis networks always affect the de modulated output The user can select whether or not the de em phasis network affects the devia tion measured and indicated by the d...

Page 6: ...pecial functions can also be used to set any measurement range or instrument function They can be used to select either of two inter nal IF s the one normally used for frequencies above 10 MHz or a narrow IF where rates and devia tions are more restricted but selectivity is increased All instru ment functions not set using these special functions remain in auto matic mode This allows the user to s...

Page 7: ...ions Besides improving the accuracy of these measure ments it greatly reduces the time involved in making measure ments Also the optional calibra tors provide a new level of modulation standard accuracy and help ensure accurate mea surements They also enhance the calibration of the modulation analyzer itself Because the modulation analyzer is useful for characterizing all types of signals in gener...

Page 8: ...nty period Hewlett Packard Company will at its option either repair or replace products which prove to be defective For warranty service or repair this product must be returned to a service facility designated by HP Buyer shall prepay shipping charges to HP and HP shall pay shipping charges duties and taxes for products returned to HP from another country HP warrants that its software and firmware...

Reviews: