prog.book : ch9sens.fm 26 Mon Feb 15 09:24:37 1999
SENSe Subsystem
[SENSe[1]]|SENSe2:CORRection:CFACtor|GAIN[1|2][:INPut][:MAGNitude]
9-26
HP E4418B/E4419B Programming Guide
[SENSe[1]]|SENSe2:CORRection:CFACtor|GAIN[1|2][:INPut][:MAGNitude] <numeric_value>
This command is used to enter a value for gain correction. The power
meter corrects every measurement by this factor to compensate for the
gain.
There are two gains which can be set:
•
CFACtor|GAIN1
is used to enter a gain correction value for the
calibration factor.
You can use either
CFACtor
or
GAIN1
in this command, both do
the same. Using
GAIN1
complies with the SCPI standard whereas
CFACtor
does not but may make your program more
understandable.
•
GAIN2
is used to enter an offset value to compensate for the power
meter setup.
Entering a value for
GAIN2
using this command automatically
turns the
[SENSe[1]]|SENSe2:CORRection:GAIN2:STATe
command to
ON
.
Gain and loss correction are a coupled system. This means that a
gain set by
[SENSe[1]]|SENSe2:CORRection:GAIN2
is
represented in the
[SENSe[1]]|SENSe2:CORRection:LOSS2?
command.
LOSS2
is coupled to
GAIN2
by the equation:
when the default unit is linear, and
when the default unit is logarithmic.
Loss
1
Gain
-------------
=
Gain
Loss
–
=
Summary of Contents for E4418B
Page 25: ...prog book chapter1 fm 1 Mon Feb 15 09 24 37 1999 1 Power Meter Remote Operation ...
Page 138: ...prog book ch2meas fm 1 Mon Feb 15 09 24 37 1999 2 MEASurement Instructions ...
Page 200: ...prog book ch3calc fm 1 Mon Feb 15 09 24 37 1999 3 CALCulate Subsystem ...
Page 228: ...prog book ch4cal fm 1 Mon Feb 15 09 24 37 1999 4 CALibration Subsystem ...
Page 242: ...prog book ch5disp fm 1 Mon Feb 15 09 24 37 1999 5 DISPlay Subsystem ...
Page 260: ...prog book ch6form fm 1 Mon Feb 15 09 24 37 1999 6 FORMat Subsystem ...
Page 264: ...prog book ch7mem fm 1 Mon Feb 15 09 24 37 1999 7 MEMory Subsystem ...
Page 294: ...prog book ch8out fm 1 Mon Feb 15 09 24 37 1999 8 OUTput Subsystem ...
Page 302: ...prog book ch9sens fm 1 Mon Feb 15 09 24 37 1999 9 SENSe Subsystem ...
Page 360: ...prog book ch10stat fm 1 Mon Feb 15 09 24 37 1999 10 STATus Subsystem ...
Page 384: ...prog book ch11syst fm 1 Mon Feb 15 09 24 37 1999 11 SYSTem Subsystem ...
Page 420: ...prog book ch12trig fm 1 Mon Feb 15 09 24 37 1999 12 TRIGger Subsystem ...
Page 434: ...prog book ch13unit fm 1 Mon Feb 15 09 24 37 1999 13 UNIT Subsystem ...
Page 442: ...prog book ch14serv fm 1 Mon Feb 15 09 24 37 1999 14 SERVice Subsystem ...
Page 452: ...prog book ch15ieee fm 1 Mon Feb 15 09 24 37 1999 15 IEEE488 2 Command Reference ...
Page 481: ...prog book progIX doc 6 Mon Feb 15 09 24 37 1999 Index 6 HP E4418B E4419B Programming Guide ...