prog.book : ch9sens.fm 7 Mon Feb 15 09:24:37 1999
SENSe Subsystem
[SENSe[1]]|SENSe2:AVERage:COUNt:AUTO <Boolean>|ONCE
HP E4418B/E4419B Programming Guide
9-7
[SENSe[1]]|SENSe2:AVERage:COUNt:AUTO <Boolean>|ONCE
This command is used to enable and disable automatic averaging.
ONCE
has no affect on the power meter.
When the auto filter mode is enabled, the power meter automatically sets
the number of readings averaged together to satisfy the averaging
requirements for most power measurements. The number of readings
averaged together depends on the resolution and the power level in which
the power meter is currently operating. Figure 9-1 lists the number of
readings averaged for each range and resolution when the power meter is
in auto filter mode.
Setting this command to
ON
automatically sets the
[SENSe[1]]|SENSe2:AVERage:STATe
command to
ON
.
Figure 9-1: Averaged Readings
If
[SENSe[1]]|SENSe2:AVERage:COUNt:AUTO
is set to
OFF
, the filter
length is set by the
[SENSe[1]]|SENSe2:AVERage:COUNt
command.
Using the
[SENSe[1]]|SENSe2:AVERage:COUNt
command
automatically disables automatic averaging.
Auto averaging is enabled by the
MEASure:POWer:AC?
and
CONFigure:POWer:AC?
commands.
10 dB
10 dB
Minimum Sensor Power
Maximum Sensor Power
P
ower Sensor
Dynamic Range
10 dB
10 dB
1
2
3
4
1
8
128
128
Resolution Setting
Number of A
verages
1
1
16
256
1
1
2
32
1
1
1
16
1
1
1
8
Summary of Contents for E4418B
Page 25: ...prog book chapter1 fm 1 Mon Feb 15 09 24 37 1999 1 Power Meter Remote Operation ...
Page 138: ...prog book ch2meas fm 1 Mon Feb 15 09 24 37 1999 2 MEASurement Instructions ...
Page 200: ...prog book ch3calc fm 1 Mon Feb 15 09 24 37 1999 3 CALCulate Subsystem ...
Page 228: ...prog book ch4cal fm 1 Mon Feb 15 09 24 37 1999 4 CALibration Subsystem ...
Page 242: ...prog book ch5disp fm 1 Mon Feb 15 09 24 37 1999 5 DISPlay Subsystem ...
Page 260: ...prog book ch6form fm 1 Mon Feb 15 09 24 37 1999 6 FORMat Subsystem ...
Page 264: ...prog book ch7mem fm 1 Mon Feb 15 09 24 37 1999 7 MEMory Subsystem ...
Page 294: ...prog book ch8out fm 1 Mon Feb 15 09 24 37 1999 8 OUTput Subsystem ...
Page 302: ...prog book ch9sens fm 1 Mon Feb 15 09 24 37 1999 9 SENSe Subsystem ...
Page 360: ...prog book ch10stat fm 1 Mon Feb 15 09 24 37 1999 10 STATus Subsystem ...
Page 384: ...prog book ch11syst fm 1 Mon Feb 15 09 24 37 1999 11 SYSTem Subsystem ...
Page 420: ...prog book ch12trig fm 1 Mon Feb 15 09 24 37 1999 12 TRIGger Subsystem ...
Page 434: ...prog book ch13unit fm 1 Mon Feb 15 09 24 37 1999 13 UNIT Subsystem ...
Page 442: ...prog book ch14serv fm 1 Mon Feb 15 09 24 37 1999 14 SERVice Subsystem ...
Page 452: ...prog book ch15ieee fm 1 Mon Feb 15 09 24 37 1999 15 IEEE488 2 Command Reference ...
Page 481: ...prog book progIX doc 6 Mon Feb 15 09 24 37 1999 Index 6 HP E4418B E4419B Programming Guide ...