Adjustment Procedure
Adjustment is performed on each channel, one at a time. Voltage output and
current output each use unique calibration constants.
Preparation
Before performing any adjustments, determine if the adjustments are to be
temporary or permanent (refer to the discussion on page 60).
If Permanent
1.
Turn off the mainframe.
2.
Remove the terminal module.
3.
Remove the module from the mainframe.
4.
Set the Cal Store Enable Jumper to the CAL position as shown below.
5.
Re-install the module in the mainframe.
6.
Make the CAL output terminal connections on the terminal module.
7.
Install the terminal module.
8.
Apply power and allow a 15 minute warm-up.
If Temporary
1.
Make the CAL output terminal connections on the terminal module.
2.
Install the terminal module.
3.
Apply power and allow a 15 minute warm-up.
Figure 5-2. Cal Store Enable Jumper
62 Adjustments
Chapter
5