Functional Verification Test
The Functional Verification Test quickly verifies that the module is
operational. Perform this test any time to verify or check the operation.
Perform this test before beginning any Quick Verification or Performance
Verification Tests.
Functional
Verification Test
Procedure
Use the following procedure to verify the module’s operation.
1.
Install the module in the mainframe and apply power.
2.
Verify the module passed its power-on self-test by executing the
SYStem:ERRor?
query and entering the results.
A passed self-test is indicated by the return
+0, “ No Error”
any other return value indicates a power-on failure.
Note
If an incorrect module address is used, the module will not respond.
Verify the module’s address before troubleshooting.
3.
Execute the full self-test by sending the
*TST?
query and entering
the result.
A passed self-test is indicated by the return
+0
Any other return value indicates a failure. Additional information
about the self-test results and self-test error codes is given in
Chapter 6 beginning on page 73.
Functional
Verification Test
Example
An example of the Functional Verification Test is shown beginning on
page 27.
Chapter 4
Verification Tests 33]
Summary of Contents for E1418A
Page 4: ...Notes 4 HP E1418A 8 16 Channel D A Converter Service Manual Contents ...
Page 8: ...Notes 8 HP E1418A D A Converter Module Service Manual ...
Page 58: ...Notes 58 Verification Tests Chapter 4 ...
Page 66: ...Figure 5 5 Internal Resistor Adjustment Connections 66 Adjustments Chapter 5 ...
Page 81: ...Figure 6 1 8 Channel Disassembly Chapter 6 Service 81 ...
Page 82: ...Figure 6 2 16 Channel Disassembly 82 Service Chapter 6 ...
Page 84: ...Notes 84 Service Chapter 6 ...
Page 90: ...Figure 7 1 HP E1418A Replaceable Parts 90 Replaceable Parts Chapter 7 ...
Page 91: ...Figure 7 2 Common Terminal Module Replaceable Parts Chapter 7 Replaceable Parts 91 ...