3-1
3
Monitoring
This chapter contains information about monitoring a patient with the CodeMaster
XL+ defibrillator/monitor. This chapter also contains details of patient preparation
that apply to the synchronized cardioversion and pacing procedures described later
in this manual.
The CodeMaster XL+ can be used for either short term or long-term cardiac
monitoring. A fully charged battery pack provides a minimum of two and a half
hours of continuous monitoring. The power cord can be connected to AC power for
unlimited monitoring periods.
3
Using Leads to Monitor
The CodeMaster XL+'s monitoring functions can be used for cardiac monitoring,
elective cardioversion, and pacing (optional). Table 3-1, “Cardiac Monitoring
Configurations,” on page 3-1 details the different ECG sources that can be used for
cardiac monitoring and monitoring applications for which each is suited.
Table 3-1
Cardiac Monitoring Configurations
For this ECG
source...
Use this cable type...
In this monitoring
application...
Electrode P/N
LEADS,
LEAD I,II,III
3-Wire:
•
6 Pin - M1731A
•
8 Pin - M1733A
•
12 Pin - M1605A/
M1500A
•
ECG Normal.
•
Synchronized
Cardioversion.
•
Pacing.
14445C
LEADS,
LEAD
I,II,III,aVR,aVL
, aVF, and V
5-Wire:
•
6 Pin - M1732A
•
8 Pin - M1734A
•
12 Pin - M1625A/
M1520A
•
ECG Normal.
•
Synchronized
Cardioversion.
•
Pacing.
40493D