AC + DC Voltage (rms responding, computed from acV, dcV)
Range
Resolution
Accuracy
DC, 20 Hz to
10 kHz
DC, 10 kHz to
30 kHz
DC, 30 kHz to
50 kHZ
DC, 50 kHz to
100 kHz
5 V
1 mV
±
(1% + 30)
±
(1.2% + 40)
±
(2.5% + 70)
±
(3.5% + 300)
50 V
10 mV
500 V
100 mV
750 V
1 V
±
(1% + 30)
DC, 20 Hz to 1 kHz
Not Specified
Measurement range:
500 mV to 500 V ranges
20 Hz to 30 kHz
5% to 100% of range
30 kHz to 100 kHz
10% to 100% of range
750 V range
75 V to 750 V
Response time: < 2 seconds for AC, 5 seconds for AC+DC on fixed range
Crest factor: <3
Common Mode Rejection Ratio (CMR) 1 k
Ω
imbalance: > 60 dB @ DC to 60 Hz
DC Current, AC Current (40 Hz to 1 kHz), 5% to 100% of range
Range
Resolution
DC Current
Accuracy
AC Current
Accuracy
Input
Resistance
Maximum
Input
500
µ
A
10 nA
±
(0.3% + 2)
±
(1% + 20)
< 1050
Ω
0.5 A (fused)
50 mA
1
µ
A
< 12
Ω
500 mA
10
µ
A
< 2.5
Ω
10 A
1 mA
±
(0.7% + 2)
< 0.05
Ω
15 A (fused)
Specifications
1-18
Summary of Contents for 974A
Page 1: ...HP 974A Multimeter User s Guide Part Number 00974 90002 March 1995 ...
Page 2: ......
Page 21: ...Adjustments ...