Step
Function
Range
Input Signal
Adjustment
(limits)
Tolerance (counts)
972A
973A
29
mA
40 mA
38.00 mA @ 100 Hz
—
±
61
±
61
30
38.00 mA @ 2 kHz
—
±
61
±
61
31
400 mA
380.0 mA @ 100 Hz
—
±
61
±
61
32
380.0 mA @ 2 kHz
—
±
61
±
61
33
10 A
10 A
10.00 A @ 100 Hz
—
±
19
±
19
34
10.00 A @ 2 kHz
—
±
19
±
19
35
Ω
400
Ω
Short
zero adjust
1
±
1
±
1
36
380.0
Ω
—
±
8
±
8
37
4 k
Ω
3.800 k
Ω
—
±
8
±
8
38
40 k
Ω
38.00 k
Ω
—
±
8
±
8
39
400 k
Ω
380.0 k
Ω
—
±
8
±
8
40
4 M
Ω
3.800 M
Ω
—
±
20
±
20
41
40 M
Ω
38.00 M
Ω
—
±
40
±
40
42
400
Ω
0
Ω
to 100
Ω
—
Tone below approx 20
Ω
43
2 V
1.000 V
—
±
12
±
12
44
100
µ
F
90.0
µ
F
9 (
±
2)
±
29
±
29
45
10
µ
F
9.00
µ
F
10 (
±
2)
±
12
±
12
46
10 nF
Open
zero adjust
1
±
1
±
1
47
100 nF
90.0 nF
11 (
±
2)
±
21
±
21
48
10 nF
9.00 nF
—
±
21
±
21
49
1000 nF
900 nF
—
±
12
±
12
50
1000
µ
F
900
µ
F
—
±
29
±
29
51
Hz (V )
4 V
9000 Hz @ 1 Vrms
—
±
2
±
2
52
Hz (A)
400
µ
A
9000 Hz @ 100
µ
A
—
±
2
±
2
1
Perform zero adjustment using
key.
Calibration Table
6 - 3