Maintenance
Operator protection from electric shock hazard is provided by a double insulated enclosure.
Refer to pages 1-4 and 1-5 for maximum voltage specifications. When servicing, use only
specified replacement parts.
Battery Replacement
Replace the battery when the symbol appears in the display or before calibration. Replace
both batteries at the same time. Use high-quality type AA alkaline (IEC LR6) batteries.
Remove the batteries if the multimeter is to be stored for extended periods of time. Refer to
the disassembly drawing on page 6-5.
Fuse Replacement
Fuse locations are shown in the diagram on page 6-5. Fuses are listed in the replaceable
part list on page 6-4. See fuse check procedure in Troubleshooting table below.
CAUTION
For continued protection use only the specified manufacturer part number or HP part number
fuse for replacement purposes.
1-15
Summary of Contents for 971A
Page 2: ...HP 971A Multimeter User s Guide Part Number 00971 90002 March 1995 ...
Page 3: ......
Page 22: ...Adjustments ...