Chapter 3, HP-IB Command Guidelines
Sequential and Overlapped Commands
Sequential and Overlapped Commands
IEEE 488.2 makes the distinction between sequential and overlapped commands.
Sequential commands complete their task before execution of the next command
can begin. Overlapped commands can run concurrently, that is, a command
following an overlapped command may begin execution while the overlapped
command is still in progress. All commands in the Test Set are sequential.
The processing architecture of the Test Set allows it to accept commands through
the HP-IB while it is executing commands already parsed into its command
buffer. While this may appear to be overlapped, commands are always executed
sequentially in the order received.
The process of executing a command can be divided into three steps:
1. Command is accepted from HP-IB and checked for proper structure and parameters.
2. Commands is sent to instrument hardware.
3. Instrument hardware fully responds after some time,
∆
t.
For example, in programming the Test Set’s RF Signal Generator it takes
< 150 ms after receipt of the frequency setting command for the output signal to
be within 100 Hz of the desired frequency. In the Test Set, commands are
considered to have “completed their task” at the end of step 2. In manual
operation all displayed measurement results take into account the instrument
hardware’s response time. When programming measurements through HP-IB the
Triggering mode selected will determine whether the instrument’s response time
is accounted for automatically or if the control program must account for it. Refer
to
"Triggering Measurements" on page 243
for a discussion of the different Trigger
modes available in the Test Set and their affect on measurement results.
Summary of Contents for 8920B
Page 10: ...10 ...
Page 16: ...Contents 16 9 Error Messages Index 673 ...
Page 17: ...17 1 Using HP IB ...
Page 48: ...48 Chapter 1 Using HP IB Remote Local Modes ...
Page 49: ...49 2 Methods For Reading Measurement Results ...
Page 61: ...61 3 HP IB Command Guidelines ...
Page 84: ...84 Chapter 3 HP IB Command Guidelines Guidelines for Operation ...
Page 85: ...85 4 HP IB Commands ...
Page 90: ...90 Adjacent Channel Power ACP ...
Page 116: ...116 Configure I O Configure ...
Page 156: ...156 Call Processing ...
Page 185: ...185 System System SYSTem ERRor Returns integer value quoted string ...
Page 242: ...242 Common Command Descriptions ...
Page 251: ...251 5 Advanced Operations ...
Page 377: ...377 7 IBASIC Controller ...
Page 458: ...458 Chapter 7 IBASIC Controller The TESTS Subsystem ...
Page 640: ...640 Chapter 8 Programming The Call Processing Subsystem Example Programs 11280 SUBEND 11290 ...
Page 643: ...643 9 Error Messages ...
Page 672: ...672 ...
Page 682: ...Index 682 ...