Triggering Measurements
Triggering Measurements
The measurement cycle is started (triggered) by the occurrence of a trigger event.
The reliability and accuracy of the measurement result, as well as the speed of the
measurement cycle are influenced by the trigger mode in effect at the time the
trigger event occurs. Some modes are faster than others; some modes provide
settling for signals that may contain transients. The best triggering mode to use
will depend upon the measurement requirements (repeatability, accuracy and
speed).
Trigger Event
The Test Set starts a measurement cycle when a valid Trigger Event is received. A
Trigger Event is analogous to telling the Test Set to “start the measurement now.”
There are three commands that can be used to issue a Trigger Event to the Test Set
through HP-IB:
•
A Group Execute Trigger Command (GET) as defined by IEEE 488.1-1987
•
A Trigger Common Command (*TRG) as defined by IEEE 488.2-1987
•
A :TRIGger:IMMediate Test Set command.
All three commands are equivalent and have the same effect when received by the
Test Set. The Test Set responds to the three commands by triggering all currently
active measurements. A measurement is defined as active if
1. it is on the currently displayed screen
2. it is in the ON state
From a programming perspective this means that the screen which contains the
measurement of interest must be made available using the DISPlay command and
that the measurement STATe must be ON.
Summary of Contents for 8920B
Page 10: ...10 ...
Page 16: ...Contents 16 9 Error Messages Index 673 ...
Page 17: ...17 1 Using HP IB ...
Page 48: ...48 Chapter 1 Using HP IB Remote Local Modes ...
Page 49: ...49 2 Methods For Reading Measurement Results ...
Page 61: ...61 3 HP IB Command Guidelines ...
Page 84: ...84 Chapter 3 HP IB Command Guidelines Guidelines for Operation ...
Page 85: ...85 4 HP IB Commands ...
Page 90: ...90 Adjacent Channel Power ACP ...
Page 116: ...116 Configure I O Configure ...
Page 156: ...156 Call Processing ...
Page 185: ...185 System System SYSTem ERRor Returns integer value quoted string ...
Page 242: ...242 Common Command Descriptions ...
Page 251: ...251 5 Advanced Operations ...
Page 377: ...377 7 IBASIC Controller ...
Page 458: ...458 Chapter 7 IBASIC Controller The TESTS Subsystem ...
Page 640: ...640 Chapter 8 Programming The Call Processing Subsystem Example Programs 11280 SUBEND 11290 ...
Page 643: ...643 9 Error Messages ...
Page 672: ...672 ...
Page 682: ...Index 682 ...