4-13
Remote Control
Performing a Measurement
Performing a Measurement
Commands to execute, stop, and abort a measurement are issued via a remote
shell software interface through the network. The commands are sent to the
test executive of the system that executes them. In addition to sending the
commands, your software can also receive the pass/fail result of a measure-
ment from the system.
When it comes to the actual performance of the tests by an operator at a com-
puter different from the system computer, a remote execution package is
invoked which communicates with a custom component of the test executive.
The test executive is initiated on the HP 86037B from the remote computer
via a version of remote shell service supported in MS Windows NT 4.0. This
remote shell service software is the supplied Denicomp Systems software.
Successful remote execution of the system includes the following steps:
1
Setup and initiation of the remote shell services software on the remote
computer.
This can be done using user-supplied remote shell software or, if it the remote
computer is a Windows NT 4.0 system, using the software located in the
C:\HPCD\ RELATED INSTALLATIONS\RSH
directory.
2
Setup and initiation of the remote shell services software on the HP 86037B. To
install this software (a product of Denicomp Systems), follow the instructions
provided in the supplied software package.
3
Initiation of part of the test executive called
MsgCenter
on the HP 86037B.
4
Initiation of the execution of a test plan, a test group, or a test.
5
Issuance of any cancel or abort commands, as necessary.
6
Display of the pass or fail results.
7
Termination of the
MsgCenter
on the HP 86037B.
Before starting the remote shells, the
.rhosts
files should be setup. The
.rhosts
file on the measurement computer should contain the IP address and user
name of the remote computer while the remote computer should have the IP
address and user name of the measurement computer. Once the remote shell
Summary of Contents for 86037B
Page 1: ...User s Guide HP 86037B Chromatic Dispersion Test System ...
Page 31: ...2 13 Operation The Main Window Figure 2 4 Repeat Measurement Selector ...
Page 72: ...3 16 Normalization Setting Normalization Preferences ...
Page 94: ...4 22 Remote Control TES Parameters Main chromatic dispersion screen ...
Page 100: ...4 28 Remote Control TES Parameters CD and Gain Measurement Settings Dialog Box ...
Page 102: ...4 30 Remote Control TES Parameters Length Measurement Settings Dialog Box ...
Page 104: ...4 32 Remote Control TES Parameters Normalization Settings Dialog Box ...
Page 106: ...4 34 Remote Control TES Parameters New Normalization Button Dialog Box ...
Page 120: ...4 48 Remote Control Output Parameters ...
Page 158: ...5 38 Maintenance Hewlett Packard Sales and Service Offices ...
Page 170: ...6 12 Installation Installation ...
Page 174: ...7 4 Reference Theory of Operation Figure 7 3 HP 86037B Block Diagram Options 121 and 122 ...
Page 175: ...7 5 Reference Theory of Operation Figure 7 4 The HP 86037B Option 114 block diagram ...
Page 190: ...7 20 Reference System Options ...
Page 191: ...8 Specifications 8 3 Regulatory Information 8 7 Specifications and Regulatory Information ...
Page 204: ...Index 6 Index ...