2-17
Operation
The Main Window
Copy Normalization wavelengths to Test wavelengths
Use this selection to copy the calibration wavelength limits to the wavelengths
used for the measurement. If the values are identical, this feature eliminates
the need to enter the wavelength limits twice.
Length Meas command
HP Defaults
This button sets the dialog box
to its default settings.
TLS Wavelength and TLS
Target Power
Edit these fields to set the
wavelength and power level of
the tunable laser source. If the specified power level is too high for the tunable
laser source, the tunable laser source supplies as much as is possible.
Average Factor
Increasing the average factor reduces noise when measuring length. Because
this factor controls the number of network analyzer sweeps which are aver-
aged to at each wavelength point, measurement time increases in proportion
to averaging factor. Noise is reduced by the reciprocal of the square root of the
averaging factor.
Sensitivity
You can select from
Normal
,
High,
and
Highest
settings. Increasing sensitivity
reduces noise but increases the measurement time. The sensitivity setting
configures the IF bandwidth used by the network analyzer.
Normal
,
High,
and
Highest
correspond to 300, 100, and 30 Hz, respectively.
Table 2-3. Example Measurement Times for 20 Measurement Steps
Test
Measurement Time
CD measurement only
2 minutes, 50 seconds
CD and Gain measurement
3 minutes, 46 seconds
Summary of Contents for 86037B
Page 1: ...User s Guide HP 86037B Chromatic Dispersion Test System ...
Page 31: ...2 13 Operation The Main Window Figure 2 4 Repeat Measurement Selector ...
Page 72: ...3 16 Normalization Setting Normalization Preferences ...
Page 94: ...4 22 Remote Control TES Parameters Main chromatic dispersion screen ...
Page 100: ...4 28 Remote Control TES Parameters CD and Gain Measurement Settings Dialog Box ...
Page 102: ...4 30 Remote Control TES Parameters Length Measurement Settings Dialog Box ...
Page 104: ...4 32 Remote Control TES Parameters Normalization Settings Dialog Box ...
Page 106: ...4 34 Remote Control TES Parameters New Normalization Button Dialog Box ...
Page 120: ...4 48 Remote Control Output Parameters ...
Page 158: ...5 38 Maintenance Hewlett Packard Sales and Service Offices ...
Page 170: ...6 12 Installation Installation ...
Page 174: ...7 4 Reference Theory of Operation Figure 7 3 HP 86037B Block Diagram Options 121 and 122 ...
Page 175: ...7 5 Reference Theory of Operation Figure 7 4 The HP 86037B Option 114 block diagram ...
Page 190: ...7 20 Reference System Options ...
Page 191: ...8 Specifications 8 3 Regulatory Information 8 7 Specifications and Regulatory Information ...