1-6
Quick Start
A Quick Tour
9
On the
File
menu, click
Save Setup As
to
save all of the settings made so far into a
setup file.
The title bar displays the name of the setup file that is currently open. You can
recall this file at any time to automatically configure the system to your spe-
cific measurement settings. You can create as many setup files as you may
need.
Normalizing the test setup
The system must warm up for two hours before test setup normalizations and
measurements can be made. While your system is warming up, spend some
time familiarizing yourself with the remaining information in this chapter.
10
Connect the device being tested to the system.
Use patch cords to connect the
SOURCE OUTPUT
and
RECEIVER INPUT
connectors
to the device you are testing. This step determines the cables required in
order to protect the system’s fiber-optic connectors from wear.
11
Remove the device being tested from the test setup and
replace the device
with a patch cord
. Retain as many of the test setup cables as possible.
12
Click
Power Monitor
, and confirm that the power levels indicate that good low-
loss connections have been made.
13
Click
New Normalization
to normalize the measurement setup. Then, click
Start
Normalization
.
describes the various normalization procedures
and adjustments available for your system.
14
After the normalization has completed, click
Close
.
Performing a measurement
15
Remove the patch cord inserted in
and reinsert the device being tested.
Be sure to use the same cables that you used in
.
16
Click
Power Monitor
, and confirm that the power levels indicate that good low-
loss connections have been made.
17
Click Verify Normalization.
Use this step to confirm that the normalization is still valid.
18
Click
Start Test
to begin a CD measurement.
Summary of Contents for 86037B
Page 1: ...User s Guide HP 86037B Chromatic Dispersion Test System ...
Page 31: ...2 13 Operation The Main Window Figure 2 4 Repeat Measurement Selector ...
Page 72: ...3 16 Normalization Setting Normalization Preferences ...
Page 94: ...4 22 Remote Control TES Parameters Main chromatic dispersion screen ...
Page 100: ...4 28 Remote Control TES Parameters CD and Gain Measurement Settings Dialog Box ...
Page 102: ...4 30 Remote Control TES Parameters Length Measurement Settings Dialog Box ...
Page 104: ...4 32 Remote Control TES Parameters Normalization Settings Dialog Box ...
Page 106: ...4 34 Remote Control TES Parameters New Normalization Button Dialog Box ...
Page 120: ...4 48 Remote Control Output Parameters ...
Page 158: ...5 38 Maintenance Hewlett Packard Sales and Service Offices ...
Page 170: ...6 12 Installation Installation ...
Page 174: ...7 4 Reference Theory of Operation Figure 7 3 HP 86037B Block Diagram Options 121 and 122 ...
Page 175: ...7 5 Reference Theory of Operation Figure 7 4 The HP 86037B Option 114 block diagram ...
Page 190: ...7 20 Reference System Options ...
Page 191: ...8 Specifications 8 3 Regulatory Information 8 7 Specifications and Regulatory Information ...
Page 204: ...Index 6 Index ...