HP 85325A Operating and Service Manual 4-3
System Performance Verification
Band-Dependent
Equipment
Table 4-2 shows equipment that is millimeter wave band dependent. Fixed
terminations, variable attenuators, and directional couplers are only usable
in one band. Therefore, if your system has more than one millimeter wave
band, you must have a termination, variable attenuator, and directional
coupler for each band.
Table 4-2
Recommended Test Equipment (banded)
Item
Critical Specification
Recommended Model or Part Number
Power Meter
No substitute
Anritsu ML83A (for V or U bands only)
Power Sensor
No substitute
R band: HP R8486A
Q band: HP Q8486A
U band: Anritsu ML715A
V band: Anritsu ML716A
W band: HP W8486A
Fixed Termination
1
Return loss >30 dB
R band: HR R910A
Q band: HP Q910A
U band: HP U910A
V band: HP V910C
W band: HP W910C
Variable Vane Attenuator
0 to 40 dB attenuation
R band: HP R382A
Q band: HP Q382A
U band: HPU382A
V band: Millitech DRA-15 (v-band)
W band: Millitech DRA-10 (w-band)
Directional Coupler
Coupling factor 10 dB
R band: HP R752C
Q band: HP Q752C
U band: HP U752C
V band: HP V752C
W band: HP W752C
1. You must have a fixed termination for each test channel of your system