![HP 85106D Manual Download Page 73](http://html.mh-extra.com/html/hp/85106d/85106d_manual_161501073.webp)
2-28
HP 85106D mm-Wave System
Operation
Example Measurements
Using Frequency List
The HP 8510C frequency list mode allows the measurement of devices at
specific user defined frequency points. In this measurement of a 94 GHz
bandpass filter, a multi-segment frequency list is used to measure the filter
transmission response from 80 GHz to 100 GHz, in 600 MHz steps, with an
increased resolution of 20 MHz from 92 GHz to 96 GHz. Figure 2-7
illustrates the A11 segments selection results. Note that either of the
segments can be displayed independently.
Figure 2-7
S
21
Response, Using Frequency List to Adjust Frequency Resolution