HP 85106D mm-Wave System
2-27
Operation
Example Measurements
Reflection
Measurements
Built in error correction provides a powerful measurement capability at
mm-wave frequencies. Systematic errors are mathematically removed from
the measurement system by measuring calibration standards with precisely
known characteristics. Figure 2-6 compares uncorrected and corrected
impedance measurements of a W-band fixed load. The slowly varying trace
shows the result using an S
11
1-port calibration.
Figure 2-6
Comparing S
11
Measurements Using Response and 1-Port Calibration