background image

Notes:

Refer to the serial number plate on rear of the chiller for model and electrical data. 

Cooling Capacity, Watts x 3.41 = BTUs/hr.  Performance specifications determined at ambient temperature of 20°C/68°F.
For 50Hz models, derate cooling capacity and pump flow by 17%.

External Pressure Reducing Assembly (for Positive Displacement Pump models) Steps down high outlet pressure to 10 to 45 psi.  Cat. No 060302

6

6000 Series Standard Chillers with Magnetic Drive Centrifugal Pump

Model

6705

6105

6155

Operating Temperature

-15° to 40°C

-15° to 40°C

+5° to 35°C

Compressor

3/4 hp

1 hp

1

1

/

hp

Cooling Capacity

@20°C  2500 Watts

@20°C  2850 Watts

@20°C  5200 Watts

@10°C  1700 Watts

@10°C  1925 Watts

@10°C  3600 Watts

@  0°C     760 Watts

@  0°C    940 Watts

@  5°C  2700 Watts

Pressure at 0 Flow Rate

10 psi

10 psi

10 psi

Flow Rate @ 0 psi

4.3 gpm / 16.3 lpm

4.3 gpm / 16.3 lpm

4.3 gpm / 16.3 lpm

Reservoir Capacity

1.65 gal / 6.25 liters

1.65 gal / 6.25 liters

6.25 gal / 23.65 liters

Dimensions       (in.)

24 x 14

3

/

8

x 26

1

/

2

24 x 14

3

/

8

x 26

1

/

2

30 x 21x 27

h x w x d            (cm)

61 x 36.5 x 67.3

61 x 36.5 x 67.3

76.2 x 53.3 x 68.6

Shipping Weight 

178 lbs

178 lbs

450 lbs

Volts, Hz, Amps

208-230V, 60Hz, 7.5A

208-230V, 60Hz, 8A

208-230V, 60Hz, 12A

Volts Range

187V to 253V

187V to 253V

187V to 253V

60Hz Cat. No.

080531

080542

080593

Volts, Hz, Amps

240V, 50Hz, 7.5A

240V, 50Hz, 8.5A

240V, 50Hz, 12A

Volts Range

198V to 264V

198V to 264V

198V to 264V

Over Voltage: Category II
50Hz Cat. No.

080536

080546

080594

Additional Specifications for "P" Series 6000 Chillers with Positive Displacement Pump

Model "P" Series

6705P

6105P

6155P

Flow Rate @ 0 psi

4 gpm / 15.2 lpm

4 gpm / 15.2 lpm

4 gpm / 15.2 lpm

Pump Pressure (Adjustable)

20 to 100psi

20 to 100psi

20 to 100psi

Shipping Weight 

178 lbs

178 lbs

450 lbs

Volts, Hz, Amps

208-230V, 60Hz, 10A

208-230V, 60Hz, 11A

208-230V, 60Hz, 12A

Volts Range

187V to 253V

187V to 253V

187V to 253V

60Hz Cat. No.

080532

080544

080596

Volts, Hz, Amps

240V, 50Hz, 10A

240V, 50Hz, 11A

240V, 50Hz, 12A

Volts Range

198V to 264V

198V to 264V

198V to 264V

Over Voltage: Category II
50Hz Cat. No.

080537

080547

080597

Additional Specifications for "T" Series 6000 Chillers with Turbine Pump

Model "T" Series

6705T

6105T

6155T

Flow Rate @ 0 PSI

3.5 gpm / 13.2 lpm 

3.5 gpm / 13.2 lpm 

3.5 gpm / 13.2 lpm

Pump Pressure (Adjustable)

20 to 100psi

20 to 100psi

20 to 100psi

Volts, Hz, Amps

208-230V, 60Hz, 9.6A

208-230V, 60Hz, 10.8A

208-230V, 60Hz, 12A

Volts Range

108V to 132V

108V to 132V

187V to 253V

60Hz Cat. No.

080533

080543

080598

Volts, Hz, Amps

240V, 50Hz, 9.6A

240V, 50Hz, 10.8A

240V, 50Hz, 12A

Volts Range

198V to 264V

198V to 264V

198V to 264V

Over Voltage: Category II
50Hz Cat. No.

080538

080548

080599

Environmental Conditions

• Indoor Use Only
• Maximum Altitude : 2000 meter

• Operating Ambient : 5° to 30°C

• Relative Humidity : 80% for temperatures to 30°C

• Pollution Degree : 2

• Class 1 : Residential, Commercial, Light Industrial

• Class 2 : Heavy Industrial

Summary of Contents for 6105P

Page 1: ...ators Manual Refrigerated Recirculating Chillers Powerful Cooling Microprocessor Controller Digital Set Read with High Flow Centrifugal Pump Positive Displacement Pump or Turbine Pump 110 116 rev D ecn 1390 ...

Page 2: ...troller Default Settings Section 4 Diagram of Models Section 5 Maintenance 5 1 Standard Magnetic Drive Centrifugal Pump Motor 5 2 Positive Displacement High Pressure Pump Motor 5 3 Fluid Filter Included With High Pressure Pump 5 4 Refrigeration Air Vents Air Filter Section 6 Troubleshooting 6 1 Unit Disabled 6 2 No Pumping 6 3 Insufficient Pumping 6 4 No Cooling or Insufficient Cooling Section 7 S...

Page 3: ... sole obligation under this warranty is limited to the repair or replacement of a defective product and the manufacturer shall not in any event be liable for any incidental or consequential damages of any kind resulting from use or possession of this product Some states do not allow A limitations on how long an implied warranty lasts or B the exclusion or limitation of incidental or consequential ...

Page 4: ...rison of Pump Options Centrifugal Pump Capacity Positive Displacement Pump Capacity 1 2 HP 1 3 HP 1 4 HP 1 HP 3 4 HP General Specifications To All Chillers Temperature Setability 0 1 C Temperature Uniformity 0 5 C Readout Selectability C or F Pump Inlet Outlet 1 2 inch NPT ...

Page 5: ...ange 198V to 264V 198V to 264V 198V to 264V Over Voltage Category II 50Hz Cat No 080506 080516 080526 Additional Specifications for P Series 6000 Chillers with Positive Displacement Pump Model P Series 6205P 6305P 6505P Flow Rate 0 psi 1 gpm 3 8 lpm 1 gpm 3 8 lpm 1 gpm 3 8 lpm Pump Pressure Adjustable 20 to 100psi 20 to 100psi 20 to 100psi Shipping Weight 160 lbs 170 lbs 170 lbs Volts Hz Amps 120V...

Page 6: ...2A Volts Range 198V to 264V 198V to 264V 198V to 264V Over Voltage Category II 50Hz Cat No 080536 080546 080594 Additional Specifications for P Series 6000 Chillers with Positive Displacement Pump Model P Series 6705P 6105P 6155P Flow Rate 0 psi 4 gpm 15 2 lpm 4 gpm 15 2 lpm 4 gpm 15 2 lpm Pump Pressure Adjustable 20 to 100psi 20 to 100psi 20 to 100psi Shipping Weight 178 lbs 178 lbs 450 lbs Volts...

Page 7: ...ibility requirements CAUSTIC CORROSIVE OR FLAMMABLE FLUIDS MUST NEVER BE USED Do not use a flammable fluid as a fire hazard may result WARNING Operation below 15 C requires antifreeze in the circulating fluid 2 3 Pump Hoses and Couplings To maintain a safe workplace and to avoid leaks special care should be taken when choosing hoses and connectors for the chiller 1 Pressure ratings Hoses should be...

Page 8: ...he pump running the reservoir s fluid level will decrease as the closed system begins to fill Add a little fluid at a time until the level in the reservoir stops going down This means that your system is filled and the air has been purged from it Replace the reservoir cap and turn it clockwise to lock it Recheck the cap for tightness Note When chillers with the standard magnetic drive centrifugal ...

Page 9: ...chiller is then fully operational Upon completion of the self test the actual fluid temperature is displayed Your desired operating temperature can now be set 2 Press the SET MENU button The degree lights flash indicating the temperature can be set or changed Turn the SETTING ADJUSTMENT knob to the desired setting This setting is accepted after pressing the SET MENU button or will be automatically...

Page 10: ...erature setability Error Screens E tF Triac alternistor failure DEF Displayed when chiller is reset to factory default settings Optional Equipment Flow or Float Switch E nF No Flow E FL Low Level Fluid 3 6 Draining the Unit Press the Power Off button and unplug the chiller Remove drain plug and reservoir cap and allow unit to drain Syphon or use a small manual pump to remove all fluid from the res...

Page 11: ...GE DRAIN DRAIN PLUG INLET Fluid In OUTLET Fluid Out REAR ACCESS PANEL TOP COVER TO ACCESS RESERVOIR TOP COVER TO ACCESS RESERVOIR PROVISION FOR 15 OR 9 PIN FEMALE CONNECTOR CIRCUIT BREAKER IEC 320 CONNECTOR LINE VOLTAGE DRAIN AIR FILTER INLET Fluid In OUTLET Fluid Out REAR ACCESS PANEL TEMPERATURE PROBE PRESSURE REGULATOR 6705 6105 6205 6305 6505 AIR FILTER ...

Page 12: ...12 6155 ...

Page 13: ...le or foreign particles a filter is included with the high pressure pump models How often the filter requires maintenance depends on the environment and use of the chiller To access this filter remove the left side panel The water filter is located behind the side brace Grip the plastic sediment bowl firmly and turn counter clockwise to remove it Clean the stainless steel wire screen found inside ...

Page 14: ......

Reviews: