Troubleshooting the MSD
Temperature symptoms
83
Mass filter (quad) heater will not heat up
•
High vacuum pump is off or has not reached normal operating conditions*
•
Incorrect temperature setpoint
•
Mass filter has not had enough time to reach temperature setpoint
•
Mass filter heater cartridge is not connected*
•
Mass filter temperature sensor is not connected*
•
Mass filter heater failed (burned out or shorted to ground)*
•
Mass filter temperature sensor failed*
•
Source power cable is not connected to the sideboard*
•
MSD electronics are not working correctly
* These will cause an error message
GC/MSD interface will not heat up
•
Incorrect setpoint
•
Setpoint entered in wrong heated zone
•
GC/MSD interface has not had enough time to reach temperature setpoint
•
GC is off
•
GC experienced a fault and needs to be reset*
•
GC/MSD interface heater/sensor cable is not connected*
•
GC/MSD heater failed (burned out)*
•
GC/MSD sensor failed*
•
GC electronics are not working correctly*
* These will cause a GC error message. GC error messages are described in the
documentation supplied with your GC.
Summary of Contents for 5973
Page 1: ...H Hardware Manual HP 5973 Mass Selective Detector ...
Page 3: ...HP 5973 Mass Selective Detector Hardware Manual ...
Page 25: ...1 Installing GC Columns To condition a capillary column 25 ...
Page 30: ...30 ...
Page 37: ...2 Operating the MSD To view MSD analyzer temperature and vacuum status 37 ...
Page 39: ...2 Operating the MSD To set monitors for MSD temperature and vacuum status 39 ...
Page 41: ...2 Operating the MSD To set the MSD analyzer temperatures 41 ...
Page 43: ...2 Operating the MSD To set the interface temperature from the ChemStation 43 ...
Page 51: ...2 Operating the MSD To remove the MSD covers 51 Upper cover Lower cover ...
Page 92: ...92 ...
Page 105: ...4 Maintaining the MSD To drain the foreline pump 105 Fill cap Drain plug Pump motor ...
Page 113: ...4 Maintaining the MSD To separate the MSD from the GC 113 ...
Page 121: ...4 Maintaining the MSD To reconnect the MSD to the GC 121 ...
Page 141: ...4 Maintaining the MSD To lubricate the side plate O ring 141 Side plate O ring ...
Page 159: ...4 Maintaining the MSD To remove a filament 159 Filament 2 Filament 1 ...
Page 161: ...4 Maintaining the MSD To reinstall a filament 161 Filament 2 Filament 1 ...
Page 210: ...210 ...
Page 214: ...214 ...
Page 232: ...232 7 Analyzer Analyzer heaters and radiators ...
Page 280: ...280 ...
Page 294: ...294 Index ...
Page 296: ...294 Chapter ...
Page 297: ...H Manual Part Number G1099 90027 Copyright 1998 Hewlett Packard Printed in USA 11 98 ...