54
2
Operating the MSD
To open the vacuum manifold
To open the vacuum manifold
Materials needed:
Gloves, clean, lint-free
large (8650-0030)
small (8650-0029)
Wrist strap, anti-static
small (9300-0969)swe
medium (9300-1257)
large (9300-0970)
1
Vent the MSD (page 52).
2
Disconnect the side board control cable and the source power cable from
the side board.
3
Loosen the side plate thumbscrews, if they are fastened.
The rear side plate thumbscrew should be unfastened during normal use. It is only
fastened during shipping. The front side plate thumbscrew should only be fas-
tened for CI operation or if hydrogen or other flammable or toxic substances are
used for carrier gas.
& $ 8 7 , 2 1
Electrostatic discharges to analyzer components are conducted to the side board where
they can damage sensitive components. Wear a grounded anti-static wrist strap
and take other anti-static precautions (see page 100) before you open the vacuum manifold.
: $ 5 1 , 1 *
7KHDQDO\]HU*&06'LQWHUIDFHDQGRWKHUFRPSRQHQWVLQWKHYDFXXPPDQLIROGRSHUDWHDW
YHU\KLJKWHPSHUDWXUHV'RQRWWRXFKDQ\SDUWXQWLO\RXDUHVXUHLWLVFRRO
& $ 8 7 , 2 1
Always wear clean gloves to prevent contamination when working in the vacuum manifold.
Summary of Contents for 5973
Page 1: ...H Hardware Manual HP 5973 Mass Selective Detector ...
Page 3: ...HP 5973 Mass Selective Detector Hardware Manual ...
Page 25: ...1 Installing GC Columns To condition a capillary column 25 ...
Page 30: ...30 ...
Page 37: ...2 Operating the MSD To view MSD analyzer temperature and vacuum status 37 ...
Page 39: ...2 Operating the MSD To set monitors for MSD temperature and vacuum status 39 ...
Page 41: ...2 Operating the MSD To set the MSD analyzer temperatures 41 ...
Page 43: ...2 Operating the MSD To set the interface temperature from the ChemStation 43 ...
Page 51: ...2 Operating the MSD To remove the MSD covers 51 Upper cover Lower cover ...
Page 92: ...92 ...
Page 105: ...4 Maintaining the MSD To drain the foreline pump 105 Fill cap Drain plug Pump motor ...
Page 113: ...4 Maintaining the MSD To separate the MSD from the GC 113 ...
Page 121: ...4 Maintaining the MSD To reconnect the MSD to the GC 121 ...
Page 141: ...4 Maintaining the MSD To lubricate the side plate O ring 141 Side plate O ring ...
Page 159: ...4 Maintaining the MSD To remove a filament 159 Filament 2 Filament 1 ...
Page 161: ...4 Maintaining the MSD To reinstall a filament 161 Filament 2 Filament 1 ...
Page 210: ...210 ...
Page 214: ...214 ...
Page 232: ...232 7 Analyzer Analyzer heaters and radiators ...
Page 280: ...280 ...
Page 294: ...294 Index ...
Page 296: ...294 Chapter ...
Page 297: ...H Manual Part Number G1099 90027 Copyright 1998 Hewlett Packard Printed in USA 11 98 ...