Operating the MSD
MSD temperatures are controlled through the data system
The MSD has independent heaters and temperature sensors for the ion
source and quadrupole mass filter. You can adjust the setpoints and view
these temperatures from the data system.
The GC/MSD interface heater is powered and controlled by the Thermal
Aux #2 heated zone of the HP 6890 Series GC. The GC/MSD interface
temperature can be set and monitored from the data system or from the
GC keypad.
Column flow is controlled through the data system
Carrier gas flow through the GC column is controlled by head pressure in
the GC. For a given head pressure, the column flow will decrease as the GC
oven temperature increases. With electronic pneumatic control (EPC) set
to
Const
Flow
(constant flow), the same column flow is be maintained regard-
less of oven temperature.
The MSD can be used to measure actual column flow. You inject a
small
amount of air or other unretained chemical, and time how long it takes to
reach the MSD. With this time, you can calculate the column flow.
The data system aids in venting
A program in the data system guides you through the venting process. It
switches off the GC and MSD heaters and the diffusion pump heater or turbo
pump at the correct time. It also lets you monitor temperatures in the MSD
and indicates when to vent the MSD.
The MSD
will
be damaged by incorrect venting. A diffusion pump will
backstream vaporized pump fluid onto the analyzer if the MSD is vented
before the diffusion pump has fully cooled. A turbo pump will be damaged
if it is vented while spinning at more than 50% of its normal operating speed.
Summary of Contents for 5973
Page 1: ...H Hardware Manual HP 5973 Mass Selective Detector ...
Page 3: ...HP 5973 Mass Selective Detector Hardware Manual ...
Page 25: ...1 Installing GC Columns To condition a capillary column 25 ...
Page 30: ...30 ...
Page 37: ...2 Operating the MSD To view MSD analyzer temperature and vacuum status 37 ...
Page 39: ...2 Operating the MSD To set monitors for MSD temperature and vacuum status 39 ...
Page 41: ...2 Operating the MSD To set the MSD analyzer temperatures 41 ...
Page 43: ...2 Operating the MSD To set the interface temperature from the ChemStation 43 ...
Page 51: ...2 Operating the MSD To remove the MSD covers 51 Upper cover Lower cover ...
Page 92: ...92 ...
Page 105: ...4 Maintaining the MSD To drain the foreline pump 105 Fill cap Drain plug Pump motor ...
Page 113: ...4 Maintaining the MSD To separate the MSD from the GC 113 ...
Page 121: ...4 Maintaining the MSD To reconnect the MSD to the GC 121 ...
Page 141: ...4 Maintaining the MSD To lubricate the side plate O ring 141 Side plate O ring ...
Page 159: ...4 Maintaining the MSD To remove a filament 159 Filament 2 Filament 1 ...
Page 161: ...4 Maintaining the MSD To reinstall a filament 161 Filament 2 Filament 1 ...
Page 210: ...210 ...
Page 214: ...214 ...
Page 232: ...232 7 Analyzer Analyzer heaters and radiators ...
Page 280: ...280 ...
Page 294: ...294 Index ...
Page 296: ...294 Chapter ...
Page 297: ...H Manual Part Number G1099 90027 Copyright 1998 Hewlett Packard Printed in USA 11 98 ...