HP5501B
Service
8-21
k.
If the sampler, photodiode, and amplifier are working properly, the oscilloscope display
will show the trace switch between two distinct levels. See Figure 8-14 below. With
continued observation these levels, which correspond to F1 and F2, will change
amplitudes since the tuning is uncontrolled. The difference between the levels varies
with time between a maximum difference and a minimum difference.
Figure 8-14. Quick Tuning Check Waveform Display
l.
Disconnect the test equipment and set HEATER jumper JMP7 to NRM. Set TEST-NORM
switch to NORM and replace
HP5501B
covers.
8-67. Testing the Liquid Crystal Switch
8-68. The following material and test equipment is required for the test:
a.
A white piece of cardboard.
b.
50 Hz Square Wave Generator (TTL Level)
c.
Test Polarizer (HP Part Number 1000-0616)
8-69. To test the liquid crystal, proceed as follows:
a.
Remove the liquid crystal from the sampler assembly as shown in Figure 8-15 below:
Figure 8-15. Liquid Crystal Removal
Summary of Contents for 5501B
Page 2: ...ii ...
Page 5: ...iii LASER HEAD SAFETY LABELS ...
Page 12: ......
Page 13: ...HP5501B Installation 1 1 Figure 1 1 HP5501B Laser Head ...
Page 21: ...HP5501B Installation 2 3 10778B LENGTH 10 METRES 32 8 FT Figure 2 1 HP 10778B Power Cable ...
Page 22: ...HP5501B Installation 2 4 10779B LENGTH 10 METRES 32 8 FT Figure 2 2 HP 10779B Reference Cable ...
Page 46: ......
Page 57: ...HP5501B Replaceable Parts 6 11 6 14 Exploded View Figure 6 1 HP5501B Laser Head Exploded View ...
Page 60: ......
Page 62: ...HP5501B Service 8 2 Figure 8 1 Schematic Diagram Notes ...
Page 66: ...HP5501B Service 8 6 Figure 8 2 HP5501B Laser Head Block Diagram ...
Page 91: ...HP5501B Service 8 31 Figure 8 18 High Voltage Power Supply Schematic Diagram ...
Page 111: ...HP5501B Service 8 51 Figure 8 23 A1 Connector Board Component Schematic Diagram ...
Page 118: ......