Chapter 4 Command Reference
[:SENSe]:FUNCtion Subtree
Programming Guide
4-29
4
COMMAND
[:SENSe]:FUNCtion:STATe? . . . <sensor_function>
This query-only command returns a Boolean value that indicates whether
the specified <sensor_function> is currently ON or OFF.
COMMENT
See
[:SENSe]:FUNCtion[:ON]
on page 4-28 for valid
<sensor_function>
strings.
[:SENSe]:POWer Subtree
This subtree controls the power-measurement function.
COMMAND
[:SENSe]:POWer:AC:REFerence . . . <numeric_value><power_units>
This command sets a reference amplitude (in dB) for display of power
measurements. It is intended for use as a measurement offset, so that a
measurement can be referenced to a known value.
RANGE
–50 dBm to 10 dBm
UNITS
dB, dBm
COMMENT
*RST: 0
RELATED
FRONT-PANEL
KEYS
Shift + Pwr Offset
COMMAND
[:SENSe]:POWer:AC:REFerence:STATe . . . <Boolean>
Determines whether amplitude is measured in absolute or relative mode.
If STATe is ON, amplitude is referenced to the value set in REFerence.
COMMENT
*RST: OFF
RELATED
FRONT-PANEL
KEYS
Power Offset On/Off
Summary of Contents for 53150A Series
Page 1: ... Programming Guide HP 53150A 151A 152A Microwave Frequency Counter ...
Page 2: ......
Page 10: ...Contents x Programming Guide ...
Page 11: ...1 Before You Start ...
Page 20: ...Chapter 1 Before You Start Related Documentation 1 10 Programming Guide 1 ...
Page 21: ...2 Command Summary A Quick Reference ...
Page 41: ...3 Programming Your Counter for Remote Operation ...
Page 101: ...4 Command Reference A Dictionary ...
Page 162: ...Chapter 4 Command Reference Common Commands 4 62 Programming Guide 4 ...
Page 163: ...5 Errors ...
Page 174: ...Chapter 5 Errors Error Types 5 12 Programming Guide 5 ...
Page 184: ......
Page 186: ... Manual Part Number 53150 90002 Printed in U S A NOVEMBER 1997 ...