Chapter 4 Command Reference
[:SENSe] Subsystem
Programming Guide
4-25
4
COMMAND
[:SENSe]:FREQuency:OFFSet . . . <numeric_value>[<frequency unit>]
Sets a reference frequency for all other absolute frequency settings in the
instrument.
RANGE
The acceptable range for the <numeric_value> parameter is 0 to 50 GHz.
UNITS
The offset frequency can be specified in Hz, KHz, or MHz only.
COMMENTS
•
This command does not affect the hardware settings of the
instrument. It affects only the entered and displayed frequencies.
•
The Counter accepts only six digits of resolution for an offset entry
through this command. For example, the command
FREQ:OFFSet 12345.678912MHz results in an offset of
approximately 12.3456 GHz.
•
The coupling equation for this command is:
Entered | Displayed frequency = (Hardware frequency ) + offset.
RELATED
FRONT-PANEL
KEYS
Freq Offset
COMMAND
[:SENSe]:FREQuency:OFFSet:STATe . . . <Boolean>
When FREQuency:OFFSet is ON, the frequency measurement results are
modified by the setting of FREQ:OFFset before being displayed or
reported (in response to a query).
RELATED
FRONT-PANEL
KEYS
Freq Offset
Summary of Contents for 53150A Series
Page 1: ... Programming Guide HP 53150A 151A 152A Microwave Frequency Counter ...
Page 2: ......
Page 10: ...Contents x Programming Guide ...
Page 11: ...1 Before You Start ...
Page 20: ...Chapter 1 Before You Start Related Documentation 1 10 Programming Guide 1 ...
Page 21: ...2 Command Summary A Quick Reference ...
Page 41: ...3 Programming Your Counter for Remote Operation ...
Page 101: ...4 Command Reference A Dictionary ...
Page 162: ...Chapter 4 Command Reference Common Commands 4 62 Programming Guide 4 ...
Page 163: ...5 Errors ...
Page 174: ...Chapter 5 Errors Error Types 5 12 Programming Guide 5 ...
Page 184: ......
Page 186: ... Manual Part Number 53150 90002 Printed in U S A NOVEMBER 1997 ...