Chapter 1 Performance Tests
Test 4: Channels 1 and 2 Frequency Accuracy (HP 53131A/132A Only)
1-28
Assembly-Level Service Guide
1
Procedure
1
Connect the equipment as shown in Figure 1-8 with the signal
generator connected to Channel 1 of the Counter.
Figure 1-8. Channels 1 and 2 Frequency Accuracy Test Setup
2
Set the Signal Generator to output 1 MHz at
−
7 dBm.
3
On the Counter, press Run key.
Verify the HP 53131A Counter reads 1 MHz ±1 mHz.
Verify the HP 53132A Counter reads 1 MHz ±200
µ
Hz.
4
Record the actual reading in the appropriate place in the
Performance Test Record (Test 4, Line 1 for HP 53131A or Test 4,
Line 3 for HP 53132A.
HP 53131A/132A
Counter
HP 8663A
Synthesized Signal Generator
N-to-BNC
Connector
Output
Channel 1
50
Ω
Feedthrough
Ref. In
(rear panel)
Input
(rear panel)
Channel 2
Summary of Contents for 53131A
Page 1: ...Assembly Level Service Guide HP 53131A 132A 181A 225 MHz Counter ...
Page 2: ......
Page 14: ...Contents xii Assembly Level Service Guide ...
Page 22: ...Preface xx Assembly Level Service Guide ...
Page 23: ...1 Performance Tests Verifying Specifications ...
Page 97: ...2 Service ...
Page 149: ...3 Replacing Assemblies Disassembly and Reassembly ...
Page 173: ...4 Retrofitting Options ...
Page 191: ...5 Replaceable Parts ...
Page 201: ...6 Backdating Manual Changes ...
Page 218: ...Chapter 6 Backdating Backdating Specifications 6 18 Assembly Level Service Guide 6 ...
Page 219: ...7 HP 53131A 132A Specifications ...
Page 235: ...8 HP 53181A Specifications ...
Page 246: ...Chapter 8 HP 53181A Specifications General Information 8 12 Assembly Level Service Guide 8 ...
Page 252: ......
Page 254: ... Manual Part Number 53131 90039 Printed in U S A MARCH 1999 ...