Chapter 1 Performance Tests
Test 2: Trigger Level (HP 53131A/132A Only)
1-20
Assembly-Level Service Guide
1
Procedure
1
Connect the output of the HP 3325B to Channel 1 of the Counter
as shown in Figure 1-6.
Figure 1-6. Trigger Level Test Setup
2
Set the HP 3325B to output a 1 MHz, 80 mVp-p square wave signal.
3
Set the HP 3325B DC OFFSET to
−
60 mV.
4
On the Counter, press Run key.
5
Now, increment the dc offset on the HP 3325B by +1 mV until the
Counter’s
Gate
annunciator flashes and continue incrementing
until the Counter displays a stable reading of approximately
1 MHz.
Observe the offset value on the display of the HP 3325B.
Record the dc offset value__________ mV.
6
Add the upper peak voltage (40 mV) of the 80 mV p-p signal to the
offset value in step 5 (For example,
−
24 mV + 40 mV = 16 mV).
This is the upper hysteresis level.
Record the upper hysteresis level__________ mV.
HP 53131A/132A
Counter
Channel 1
HP 3325B
Synthesizer
Output
Channel 2
Summary of Contents for 53131A
Page 1: ...Assembly Level Service Guide HP 53131A 132A 181A 225 MHz Counter ...
Page 2: ......
Page 14: ...Contents xii Assembly Level Service Guide ...
Page 22: ...Preface xx Assembly Level Service Guide ...
Page 23: ...1 Performance Tests Verifying Specifications ...
Page 97: ...2 Service ...
Page 149: ...3 Replacing Assemblies Disassembly and Reassembly ...
Page 173: ...4 Retrofitting Options ...
Page 191: ...5 Replaceable Parts ...
Page 201: ...6 Backdating Manual Changes ...
Page 218: ...Chapter 6 Backdating Backdating Specifications 6 18 Assembly Level Service Guide 6 ...
Page 219: ...7 HP 53131A 132A Specifications ...
Page 235: ...8 HP 53181A Specifications ...
Page 246: ...Chapter 8 HP 53181A Specifications General Information 8 12 Assembly Level Service Guide 8 ...
Page 252: ......
Page 254: ... Manual Part Number 53131 90039 Printed in U S A MARCH 1999 ...