Chapter 6 Backdating
The HP 53131A Calibration Procedures
Assembly-Level Service Guide
6-13
6
3
Connect the HP 8130A Pulse Generator output to Channel 1 of the
Counter as shown in Figure 2-3.
CHANGE 1 (Firmware Revisions 3317, 3335, 3402, 3427)
(Continued)
Equipment
HP 8130A Pulse Generator (or equivalent)
HP 10503A BNC Cable
Figure 2-3. TI Calibration Setup
4
Set the HP 8130A as follows:
PERIOD: 100 ns
WIDTH: 50 ns
High: 0.50 VLow :
−
0.50 V
Input Mode: Normal
NOTE
Any external timebase connected to the Counter is ignored during this
calibration.
5
Press Enter key.
The Counter displays
CALIBRATING
, and then it should display
TI CAL PASS
.
If the fail message is displayed, refer to the troubleshooting section in this
chapter.
HP53131A
Counter
Channel 1
HP 8130A
Pulse Generator
Output
Summary of Contents for 53131A
Page 1: ...Assembly Level Service Guide HP 53131A 132A 181A 225 MHz Counter ...
Page 2: ......
Page 14: ...Contents xii Assembly Level Service Guide ...
Page 22: ...Preface xx Assembly Level Service Guide ...
Page 23: ...1 Performance Tests Verifying Specifications ...
Page 97: ...2 Service ...
Page 149: ...3 Replacing Assemblies Disassembly and Reassembly ...
Page 173: ...4 Retrofitting Options ...
Page 191: ...5 Replaceable Parts ...
Page 201: ...6 Backdating Manual Changes ...
Page 218: ...Chapter 6 Backdating Backdating Specifications 6 18 Assembly Level Service Guide 6 ...
Page 219: ...7 HP 53131A 132A Specifications ...
Page 235: ...8 HP 53181A Specifications ...
Page 246: ...Chapter 8 HP 53181A Specifications General Information 8 12 Assembly Level Service Guide 8 ...
Page 252: ......
Page 254: ... Manual Part Number 53131 90039 Printed in U S A MARCH 1999 ...