Model 400F /FL
Section
I
SECTION
GENERAL
1-1. DESCRIPTION.
1-2. The �hp- Models 400F and 400FL are versatile
ac
voltmeters
and dB
meters. Both models can be
used as wideband amplifiers. The Model 400F is
primarily intended for voltage measu rements, where
as the Model 400FL is primarily a dB meter.
How
ever. both meters indicate both volts and dB. The
40DF has
a
linear
ac
scale with a logarithmic dB
scale underneath. and the 400FL has a linear dB
scale with a logarithmic ac scale underneath. Since
the difference in scales is the only difference between
the two instruments. this manual will use the term
400F /FL in reference to both instruments.
1-3. Figure 1-1 shows both the Model 400F and the
Mode1400FL. Table 1-1 is a list
of
specifications.
1-4. OPTION (400F ONLY).
1-5. Option 01 is a standard -hp_ Model 400F AC
Voltmeter which has a dB scale that reads from -15
INFORMATION
to +2 instead
of
from -12 to +2. The dB scale is
placed at the top of the meter face for better
resolution.
1-6. INSTRUMENT AND MANUAL
IDEN TIFATION.
1-7. Hewlett-Packard instruments are identified by
a two-section, eight-digit serial number (000-00000).
If
the first three digits of the serial number on your
instrument do
not
agree with those on the title page
of
this manual, change sheets supplied with the man
ual will define differences between your instrument
and the Model 400F /FL described in this manual.
1-8.
If
a letter prefixes the serial number, the
instrument was manufactured outside the United States.
1-9. BACKDATIN G IN FORMATION.
1-10. Appendix C contains backdating information
that adapts this manual to instruments with serials
prefixed 617.
Table 1-1. Specifications (Cont'd)
MODEL 400F
Accuracy: ±
(%
Full Scale +
%
Reading)
300 �V TO 300 V RANGES
Frequency
20 Hz
40
Hz
100 Hz
1
MHz
2 MHz
4 MRz
±
(2
+
2 )
±(1 + 1 )
±(1/2 + 1/2)
±Cl
+ 1 )
± (2
+
2)
100 �V RANGE
Frequency
30 Hz
60
Hz
±
(2
+
2)
±(1 + 1 )
MODEL 400FL
Accuracy: ±
%
Reading
300 � V TO 300 V RANGES
Frequency
20 Hz
40 Hz
100 Hz
±4
±2
±1
100 IJ,V RANGE
Frequency
30 Hz
60 Hz
±4
±2
01795-2
100 kHz
±1
1 MRz
100 kHz
+1 -8
500 kHz
+0
-7
2 MHz
500 kHz
4 MHz
±4
1-1
Summary of Contents for 400F
Page 2: ...Verso Filler Page ...
Page 3: ...400F FL AC VOLTMETER OPERATING AND SERVICE MANUAL HEWLETT PACKARD ...
Page 6: ...Verso Filler Page ...
Page 16: ...Section IV Figure 4 1 4 0 O 1 NO 0 0 ModeI400F FL u 0 8 0 j JO W t 01795 1 ...
Page 28: ...Verso Filler Page ...
Page 33: ...Verso Filler Page ...
Page 53: ......
Page 54: ...HEWLETTitPACKARD 00467 90003 Printed in U SA ...