TM 11-6625-1514-15
Section V
Paragraphs 5-1 to 5-8
SECTION V
MAINTENANCE
5-1. SCOPE.
procedures. If an adjustment or replacement of parts
5-2. This section contains complete instructions for
is made without following instructions or understanding
the effects, further trouble shooting may be complicated.
repairing and calibrating the voltmeter. This material
is covered in the following groups of paragraphs:
b. Do not remove tubes when the voltmeter is turned
on. Before replacing tubes refer to paragraph 5-10.
Lead
Paragraph
Topic
5-5. TEST EQUIPMENT REQUIRED.
Precautions
5-6. The test equipment required for complete testing
Test Equipment Required
of the voltmeter is listed in figure 5-1. Equivalent
Meter Zero Adjustment
instruments may be substituted for those listed.
Cabinet Removal
Tube Replacement
5-7. METER ZERO ADJUSTMENT.
Replacement of Special Parts
Trouble Shooting
5-8. The meter is properly zero-set when its pointer
Testing the Power Supply
rests over the zero calibration mark on the meter scale
Testing Voltmeter Performance
when the instrument is 1) at normal operating tempera-
Calibration and Frequency Response
ture, 2) in its normal operating position, and 3) turned
Adjustments
off. Adjust the zero-set if necessary, as follows:
5-3. PRECAUTIONS.
a. Allow the voltmeter to operate for 20 minutes so
that the meter movement will reach normal operating
5-4. Observe the following precautions:
temperature.
a. Make no adjustments and replace no parts in the
b. Turn the voltmeter off and allow one minute for all
voltmeter except as described in one of the following
capacitors to discharge.
INSTRUMENT TYPE
REQUIRED CHARACTERISTICS
USE
DESIGNATION
Electronic
0 to 300 a-c and d-c volts;
Voltage and resistance
ME -26 B/U or
Multimeter
accuracy of ±3% or better;
measurement.
H-P 410B
input impedance 100 megohms.
Oscillator
10 cps to 300 kc; 3 volts
Signal source for
output into 50-ohm load.
testing and calibration
H-P 200S
Voltmeter Calibrator
400-cps output voltage;
Calibrating voltmeter
(Precision Voltage
0.001 to 300 volts
at mid-frequencies.
H-P 738BR
Source)
in 10-db steps ±0.2%; 0.1 to
1.0 volt in 0.1 volt steps ±0.2%.
Frequency
300-kc to 4-mc range;
Calibrating voltmeter
Response
3 volts output into 50-ohm load;
frequency response.
H-P 739A
Test Set
10-db steps, 0 to 70 db.
Oscilloscope or
10-cps to 4-mc range.
Trouble shooting by
AC Voltmeter
H-P 160B or
signal tracing.
H-P 400D
Variable
Adjust line voltage between 103
Checking voltmeter
Transformer
and 127V ac with 1-amp load.
operation with
CN-16/U or
varying line
Ohmite VT2
voltage.
D-C Current
Clip-on type measurement;
Checking load on
Test Set
current range up to 100 ma.
power supply.
H-P 428B
(Milliammeter)
Figure 5-1. Test Equipment Required
00102-3
5-1
Summary of Contents for 4001
Page 4: ......
Page 6: ......
Page 9: ......
Page 10: ...Figure 1 1 TM 11 6625 1514 15 1 0 Change 1 ...
Page 12: ......
Page 21: ...TM 11 6625 1514 15 Section III Figure 3 4 Impedance Correction Graph 3 5 ...
Page 22: ...Figure 4 1 TM 11 6625 1514 15 Section IV 4 0 00102 2 ...
Page 26: ......
Page 30: ...TM 11 6625 1514 15 Section V Figure 5 3 Left Side View of Voltmeter Chassis 5 4 00102 3 ...
Page 36: ...Figure 5 8 TM 11 6625 1514 15 Section V 5 10 00102 3 ...
Page 37: ...TM 11 6625 1514 15 Section V Figure 5 9 Diagram of RANGE Switch 5 11 ...
Page 38: ......
Page 39: ...Figure 5 10 TM 11 6625 1514 15 Section V 00102 3 5 13 ...
Page 40: ......
Page 58: ...TM 11 6625 1514 15 Section VIII Numerical Index HEWLETT PACKARD STOCK NO INDEX 8 2 00102 3 ...
Page 60: ...TM 11 6625 1514 15 Section IX Reference Designation Index 9 2 00102 1 ...
Page 61: ...TM 11 6625 1514 15 Section IX Reference Designation Index 00102 1 9 3 ...
Page 62: ......
Page 68: ......
Page 70: ......
Page 71: ......
Page 72: ...PIN 019300 000 ...
Page 73: ......